Performance tests
54
10. In the Setup tab, enable the channel's output.
11. Click the Play button on-screen or press the button on the front panel of the
generator.
12. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
13. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1.
14. Use the following formula to compensate the voltage for the 50 Ω BNC
termination:
V_high = [(Term_R + 50) / (2×Term_R)] × Measured_voltage_1
Where Term_R is the resistance of the 50 Ω BNC termination. (See page 44,
Termination resistance measurement.) procedure.
15. Verify that the voltage difference |(V_high – V_low)| falls within the limits
given in the table. (See Table 36 on page 53.)
16. Repeat steps 9 through 15 for each bias setting in the table. (See Table 36
on page 53.)
17. Repeat steps 9 through 16 until all channels are checked, modifying the
instructions with channel number under test.
18. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
NOTE.
This is the start of testing the optional AC Amplified output path.
If the optional AC Amplified output is licensed, continue with this procedure.
If not, skip to step 21.