Tektronix AWG5200 Series Technical Reference page 50

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

Brief procedures
38
o. Set the Frequency to 5 GHz.
p. Click the Compile Settings icon to open the compile settings dialog screen.
q. In the Name field, change the name to Waveform_5 GHz.
r. Close the compile settings dialog screen.
s. Click Compile.
6. Set the spectrum analyzer as follows:
a. Press the Preset button to set the analyzer to its default settings.
b. Display the Spectrum measurement.
c. Set Center Frequency to 1 GHz.
7. Click the Setup tab on the generator display.
a. Change the Output Path to AC Direct.
8. Click the Home tab on the display.
9. Enable the channel's output.
10. In the Waveform List window, assign the Waveform_1 GHz waveform to
the channel under test.
11. Press the Play button, or click Play on the display.
12. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
13. Check that the waveform is properly displayed on the signal analyzer screen.
AWG5200 Series Technical Reference