Table 7: Signal output characteristics (cont.)
Characteristics
AC Direct
AC Amplified
(Option AC)
Amplitude adjustment resolution
Output paths
DC High BW
DC High Volt
(Option HV)
AC Direct
AC Amplified
(Option AC)
DC amplitude accuracy
Output path
DC High BW
DC High Volt
(Option HV)
AC amplitude accuracy
Output path
AC Direct
AC Amplified
(Option AC)
DC Offset range
Output path
DC High BW
DC High Volt
(Option HV)
DC Offset resolution
Output path
DC High BW
DC High Volt
(Option HV)
AWG5200 Series Technical Reference
Description
–17 dBm to –5 dBm.
10 MHz to 3.5 GHz.
–85 dBm to 10 dBm (10 MHz to 3.5 GHz.)
–50 dBm to 10 dBm (3.5 GHz to 5 GHz.)
Amplitude accuracy and flatness degrades at frequencies beyond 3.5 GHz and below –50 dBm
output amplitude. It is not recommended to operate in this region.
1.1 mV or 0.1 dB.
1.1 mV or 0.1 dB.
0.1 dB
0.1 dB
Within ±5 °C of internal self calibration temperature.
Amplitude < 100 mV: ±(5% of amplitude).
Amplitude 100 mV to 750 mV: ±(2% of amplitude).
Amplitude 100 mV to 1.5 V (Option DC): ±(2% of amplitude).
Amplitude < 160 mV: ±(5% of amplitude).
Amplitude 160 mV to 5 V: ±(2% of amplitude).
0.5 dB at 100 MHz (0 °C to 45 °C)
1 dB at 100 MHz (45 °C to 50 °C)
0.5 dB at 100 MHz (0 °C to 45 °C)
1 dB at 100 MHz (45 °C to 50 °C)
± 2 V into 50 Ω to ground.
± 4 V into high resistance or matching voltage termination.
± 2 V into 50 Ω to ground.
± 4 V into high resistance or matching voltage termination.
1 mV
1 mV