Figure 4-10: Amplitude Accuracy Initial Test Hookup - Tektronix AWG510 Service Manual

Arbitrary waveform generator
Table of Contents

Advertisement

Performance Verification
Amplitude and Offset Accuracy Tests (Normal Out)
Check Amplitude
Accuracy
4-32
These procedures check the accuracy of the AWG500–Series Waveform
Generator normal waveform outputs; amplitude and offset.
NOTE. The amplitude and offset accuracy checks are structured as a continuous
test. After Check Amplitude Accuracy, the next test uses the control settings from
the last test and uses the next step in the sequence file.
A 50 Ω coaxial cable, a 50 Ω precision terminator, a BNC (female) to
Equipment
required
dual banana adapter, and a digital multimeter (DMM).
Prerequisites
The instrument must meet the prerequisites listed on page 4-13.
1. Install the test hookup and set test equipment controls:
a. Hook up DMM: Connect the AWG500–Series Waveform Generator CH1
output through a 50 Ω BNC coaxial cable, a 50 Ω precision terminator,
and an adapter (BNC-to-dual banana) to the DMM input connector (see
Figure 4–10).
AWG500 Series Waveform Generator
50 W coaxial cable

Figure 4-10: Amplitude accuracy initial test hookup

b. Set the DMM controls:
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . Auto
Input . . . . . . . . . . . . . . . . . . . . . . . . . . Front
2. Set AWG500–Series Waveform Generator controls and select the sequence
file:
Dual Banana
Adapter
50 W
Terminator
DMM
AWG510 & AWG520 Service Manual