Tektronix AWG5200 Series Technical Reference page 46

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

Brief procedures
34
e. CH 1 through CH 4 position: adjust as necessary to display all four traces
f.
Trigger source: CH1
g. Trigger level: 0 mV
h. Trigger slope: Positive
i.
Trigger mode: Auto
4. Click the Home tab on the display.
5. Click the Reset to Default Setup button in the toolbar.
6. From the Waveform List window, assign the waveform PV_Square.wfmx to
the channel under test.
7. In the Setup tab, under Output Settings, set the Resolution to 12+4 Mkrs.
8. Enable the channel's output.
9. Click the Play button on-screen or on the front panel.
10. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
11. Check that the CH1:1 through CH1:4 waveforms are properly displayed on
the test oscilloscope screen.
AWG5200 Series Technical Reference