Tektronix AWG5200 Series Technical Reference page 43

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

Checking the analog
channel outputs
AWG5200 Series Technical Reference
Required equipment
Oscilloscope
One TCA-SMA adapter
One 50 Ω SMA cable
One 50 Ω SMA termination
1. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
2. Connect channel 1 (+) of the generator to channel 1 of the test oscilloscope
using a 50 Ω SMA cable and a TCA-SMA adapter.
3. Connect channel 1 (–) of the generator to channel 2 of the test oscilloscope
using a 50 Ω SMA cable and a TCA-SMA adapter.
4. Set the test oscilloscope as follows:
a. Vertical scale: 200 mV/div (CH 1 and CH 2)
b. Horizontal scale: 100 ns/div
c. Input coupling: DC (CH 1 and CH 2)
d. Input impedance: 50 Ω (CH 1 and CH 2)
e. CH 1 and CH 2 position: +2 div (if necessary)
f.
Trigger source: CH1
g. Trigger level: 0 mV
h. Trigger slope: Positive
i.
Trigger mode: Auto
5. Click the Home tab on the display.
Brief procedures
Prerequisites
None
31