Tektronix AWG5200 Series Technical Reference page 63

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

AWG5200 Series Technical Reference
12. Click the Play button on-screen or press the button on the front panel of the
generator.
13. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
14. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1.
15. Use the following formula to compensate the voltage for the 50 Ω BNC
termination:
V_high = [(Term_R + 50) / (2×Term_R)] × Measured_voltage_1
Where Term_R is the resistance of the 50 Ω BNC termination. (See page 44,
Termination resistance measurement.) procedure.
16. Verify that the voltage difference |(V_high – V_low)| falls within the limits
given in the table. (See Table 35 on page 50.)
17. Repeat steps 10 through 16 for each offset setting in the table. (See Table 35
on page 50.)
18. Move the SMA-BNC adapter from the (+) connector to the (–) connector and
move the 50 Ω SMA termination from the (–) connector to the (+) connector
of the channel under test.
19. Repeat steps 10 through 17 for the (–) output.
20. Repeat steps 10 through 19 until all channels are checked, modifying the
instructions with channel number under test.
Performance tests
51