Weekly Cleaning And Check Procedures - Kodak X-Omat 480 RA Service Manual

Hide thumbs Also See for X-Omat 480 RA:
Table of Contents

Advertisement

[2]
Use a clean cloth to wipe all chemical residue from the processing section of the processor. To prevent
contamination, do not use the same cloth for the FIXER and DEVELOPER sections.
[3]
Be sure that the TOP COVER remains open approximately 2 inches (5.1 cm) when the processor is
deenergized.

Weekly Cleaning and Check Procedures

For these procedures the processor must be deenergized. See page 1-7 if necessary.
To prevent contamination of the DEVELOPER and FIXER SOLUTIONS when you remove the
FIXER RACK, place the SPLASH GUARD between the DEVELOPER and FIXER TANKS. Use
the RACK DRIP TRAY when you remove or install any of the RACKS.
[1]
Remove the EVAPORATION COVERS, all CROSSOVER ASSEMBLIES, and all RACKS.
[2]
Clean the COVER on the 5600 CIRCUIT BOARD and the opening at the FILM FEED TRAY.
[3]
Rinse and wipe the removed parts with a damp cloth.
[4]
Clean the DETECTOR CROSSOVER with a soft fiber brush and warm water. Allow the RACK to air dry.
[5]
Check that all ROLLERS on all RACKS rotate freely.
Be sure that the SPLASH GUARD is
installed between the TANKS. Install
the RACKS slowly to prevent the
splashing of chemicals.
[6]
Install the RACKS, CROSSOVER
ASSEMBLIES, and EVAPORATION
COVERS. Check that each assembly seats
firmly.
[7]
Clean the REPLENISHER STRAINERS
located between the REPLENISHMENT
TANKS and the PUMPS.
[8]
Check that the SLOTS in the DRYER AIR
TUBES are clean and oriented correctly.
246616
IMPORTANT
SPLASH
GUARD
Figure 7-2 Splash Guard and Drip Tray
RACK
DRIP
H108_0035ACA
TRAY
H108_0035AA
7-3