Final Checkout
Do not place the FILM TOOL between the DETECTOR ROLLERS during the next step.
[16]
Check that the ENTRANCE DETECTOR SWITCH is adjusted correctly:
(a)
Insert the 43.18 cm (17 in.) end of the FILM TOOL under the FILM ACCUMULATOR LEDs.
(b)
If the REPLENISHER PUMPS begin operating after approximately 12 seconds, one or both of the
ENTRANCE DETECTOR SWITCHES is not adjusted correctly. Repeat steps 9-16 of the
adjustment procedure.
(c)
If the REPLENISHER PUMPS do not begin operating, guide the 4-inch FILM STRIP, cut from the
FILM TOOL, into the center of the DETECTOR ROLLERS and allow it to pass through the
processor 3 times. The REPLENISHER PUMPS should begin operating after the FILM TOOL has
been inserted 3 times.
(d)
If the REPLENISHER PUMPS do not begin operating after the FILM TOOL has been inserted 3
times, do steps 9-16 of the adjustment procedure.
[17]
Set the processor to the previous cycle.
[18]
Remove the JUMPER and connect CONNECTOR J29.
246616
DETECTOR ROLLERS
ENTRANCE DETECTOR SWITCH
Figure 6-6 Guiding the Film Strip into the Detector Rollers
4-inch FILM STRIP
H108_0167BCA
H108_0167BC
6-7