[12]
Adjust the SWITCH PLATE ASSEMBLY:
(a)
Loosen the 2 SCREWS.
(b)
Move the SWITCH PLATE ASSEMBLY down, approximately 2.5 mm (0.100 in.).
(c)
Wait 12 seconds.
(d)
If the REPLENISHER PUMPS begin operating, tighten the 2 SCREWS and continue with step 13.
(e)
If the REPLENISHER PUMPS do not begin operating, repeat step 12.
[13]
Remove the adhesive tape and pull the FILM TOOL away from the DETECTOR ROLLERS. The
REPLENISHER PUMPS will stop.
The time required to stop the REPLENISHER PUMPS corresponds to the amount of time the
FILM TOOL remained under the FILM ACCUMULATOR LEDs.
[14]
Release the NON-DRIVE SIDE MAGNET SWING ARM.
[15]
Do steps 7-14 again to adjust the NON-DRIVE SIDE ENTRANCE DETECTOR SWITCH.
NON-DRIVE SIDE
MAGNET
SWING
ARM
RESTRAINER
DETECTOR
ROLLER
6-6
NOTE
Figure 6-5 Adjusting the Switch Plate Assembly
SCREW (2)
SWITCH PLATE
ASSEMBLY
FILM ACCUMULATOR LEDs
REPLENISHMENT PUMPS
H108_0166DCB