Use unprocessed film to make a 2.5 cm (1.0 in.) x 12.5 cm (5.0 in.) FILM TOOL.
[5]
Connect the portable computer to PIC CONNECTOR P204. Use CABLE TL-4391. Allow the portable
computer to establish communications with the processor.
[6]
From the ''Specific Test Mode'' main menu, select the ''Processor Sensor Test'' option. If necessary, see
the ''User Manual for Kodak X-Omat RA Processors Software Diagnostics'' Publication No. 636719.
[7]
At the feed end of the processor, insert the FILM TOOL into the DETECTOR CROSSOVER ASSEMBLY.
Align the edge of the FILM TOOL with the drive side of the FEED TRAY.
[8]
Check that ''Film detected'' is indicated on the
screen, as the FILM TOOL is inserted.
[9]
Remove the FILM TOOL from the
DETECTOR CROSSOVER ASSEMBLY.
[10]
Check that ''No film detected'' is indicated on
the screen.
[11]
If the indications in steps 8 and 10 do not
occur, adjust the position of the DRIVE-SIDE
SWITCH PLATE ASSEMBLY and/or the
ACTUATOR MAGNET. See Figure 6-7.
[12]
Do steps 7 - 11 until the correct indications
occur when the FILM TOOL is inserted into
and removed from the DETECTOR
CROSSOVER ASSEMBLY.
WASSEREINLAUFSCHLAUCH
ENTWICKLER-/
..
FIXIERUBERHEBER
EINZUGSROLLEN-/
..
UBERHEBER-EINHEIT
246616
Figure 6-8 Inserting the Film Tool
PROCESSOR SENSOR TEST
Dev Tank Sensor
:
Fix Tank Sensor
:
Film Detect Sensor
: Film detected
Cover Switch
:
Any key=Continue
PROCESSOR SENSOR TEST
Dev Tank Sensor
:
Fix Tank Sensor
:
Film Detect Sensor
: No film detected
Cover Switch
:
Any key=Continue