Technical Information
Table 12 Range Stop/Sensitivity Requirements
Range Stop
OFF
-11
-10
-9
-8

Analog Output

Analog Output
For added accuracy add .06 to V
NOTE
Leak Rate = 10
Bar Graph
Agilent Helium Leak Detector User Manual
Max Gain
6.9
6.9
10
20
50
Configures the analog outputs:
Linear 10 V/DEC
Linear output running from 0 to 10 V for each decade.
The following figure shows the linear output voltage
as a function of leak rate.
Log 1 V/DEC
Log output at 1 V per decade. The figure 59 shows the log
output voltage as a function of pressure.
out
Vout+ .06
11
/10
.
Configures the bar graph:
Linear Only
The linear bar graph leak rate appears across the top
of all screens when selected. In the Linear Bar Graph
display mode, the bar graph display represents the
mantissa of the leak rate, spanning from 0 to 10. The
numerical display (located to the upper right of the bar
graph) indicates the leak rate exponent value. This
mode offers exceptional resolution within a decade.
MDL
5.00E-12
ON at
5.00E-12
5.00E-11
5.00E-10
5.00E-09
117