I 2 of 5
Format: <Krstatus,check digit status,check digit output status,code length
#1,code length #2>
status:
check digit status:
0 = Disabled
0 = Disabled
1 = Enabled
1 = Enabled
code length #1:
code length #2:
Zero or any even
Zero or any even
number from 2 to 30.
number from 2 to 30.
Default is 10.
Default is 6.
Example: To set Fixed Code Length #1 to 8 and Fixed Code Length #2 to 4,
enter <Kr,,,8,4> or <Kr1,0,0,8,4>.
UPC/EAN
Format: <KsUPC status,EAN status,supplementals status,separator
/
status,separator character>
UPC status:
EAN status:
0 = Disabled
0 = Disabled
1 = Enabled
1 = Enabled
(UPC must also be
enabled)
separator status:
separator character:
0 = Disabled
Any ASCII character
(except NUL). Default is
1 = Enabled
a comma (,).
Example: To change Supplementals to Required, and change Separator
Character to a dash (–), enter <Ks1,1,2,1,–> or <Ks,,2,1,–>.
Code 128
Format: <Ktstatus,fixed code
length status,code length>
status:
fixed code length status:
0 = Disabled
0 = Disabled
1 = Enabled
1 = Enabled
Example: To enable Code 128, enable Fixed Code Length, and set Code
Length to 9, enter <Kt1,1,9>.
MS-610 Scanner User's Manual
Code Types Commands
check digit output status:
0 = Disabled
1 = Enabled
supplementals status:
0 = Disabled
1 = Enabled
2 = Required
code length:
Any number from 1 to 31.
Default is 10.