Checking The Developer Temperature; Checking The Film Transport - Kodak RP X-OMAT M6B Installation Instructions Manual

Hide thumbs Also See for RP X-OMAT M6B:
Table of Contents

Advertisement

Checking the Developer Temperature

[1]
When the DEVELOPER LAMP begins to flash, remove the TOP COVER.
[2]
Insert a THERMOMETER of known accuracy into the DEVELOPER TANK.
[3]
Install the TOP COVER.
[4]
Operate the processor for 15 minutes to stabilize the temperature.
[5]
Check that the temperature on the DEVELOPER TEMPERATURE METER is the same as on the
THERMOMETER (35°C [95°F] set at the factory).
THERMOMETER
[6]
Check that the following are installed:
• RACKS
• CROSSOVERS
• COVERS
[7]
Check that the LIGHT-TIGHT GASKET is tight against the wall.

Checking the Film Transport

[1]
Feed three 35 x 43 cm (14 x 17 in.) films into the processor. Check that the RACKS move the film
through the processor correctly.
246626
See the Service Manual for the adjustment procedure.
Figure 46 Placing the Thermometer
NOTE
H048_0098BCA
H048_0098BA
47