Connecting the Exhaust
[1]
Connect 7.6 cm (3 in.) ELBOWS and rigid DUCT or flexible EXHAUST HOSE between the processor
EXHAUST PORT and the BUILDING EXHAUST DUCT.
Do not make a solid connection at the BUILDING END. Do not connect the EXHAUST DUCT to
the processor at this time.
[2]
Measure the static pressure in the DUCT using a modified J TUBE (CHECK TUBE 592380) and AIR
METER TL-2431. Make the measurement 30.5 cm (12 in.) from the end of the DUCT to be connected
to the processor.
32
NOTE
AIR FLOW
30 cm
(12 in.)
AIR METER
Figure 33 Measuring the Static Pressure
EXHAUST HOSE
RUBBER HOSE or
MODIFIED J TUBE
TAPE (3 places)
HOSE SUPPORT
CENTER CONNECTOR
H048_0118BCA
H048_0118BA