WEEKLY CARE AND CLEANING OF THE KINAIR MEDSURG PULSE
SYSTEM
In addition to daily care and cleaning, the KinAir MedSurg Pulse Cover Sheet should be replaced weekly or more often, as
needed. Excessively soiled mattress components and Cover Sheets should be replaced and laundered. Contact your
ArjoHuntleigh representative for local Infection Control procedures.

PREVENTIVE MAINTENANCE

NOTE: It is recommended that all preventative and battery maintenance procedures be performed only by qualified
ArjoHuntleigh service personnel or ArjoHuntleigh-approved personnel.
Monthly Cleaning - The cushions should be laundered at least monthly.
Air Filter - Clean / change Air Filter(s), when necessary.
Battery Inspection - Inspect batteries and replace in pairs, if batteries are more than 28 months old from the date of
manufacture. Install a Battery History label on the battery cover with the recorded date when the batteries should next be
replaced. Apply label in a visible location so that all information is displayed.
NOTE: When not in service, bed must remain plugged in to maintain battery charge.