Heater Menu; Alarm Silence - Arjohuntleigh KinAir MedSurg Pulse User Manual

Table of Contents

Advertisement

Turn Heater On or Off and adjust its target temperature.
Raise and lower head, foot sections, or the entire patient support surface.

Heater Menu

Use the Heater Menu to view current status and adjust heater settings:
Press HEAT ON / OFF to activate Heater.
Press °F / °C button to change temperature from degrees Celsius to degrees
Fahrenheit or from degrees Fahrenheit to degrees Celsius.
Press TEMP UP or TEMP DOWN arrows to adjust target Heater setting.

Alarm Silence

Use the ALARM SILENCE button to silence the following alarm conditions:
NOTE: An ALARM SILENCE button is also located on the Auxiliary Control Panel.
SEAT DEFLATE - Alarm will sound when Seat Deflate has been activated for
longer than 12.50 minutes. If more time is required, pressing ALARM SILENCE
provides another 12.50 minutes of silence.
INSTAFLATE - Alarm will sound when InstaFlate is engaged. Pressing ALARM SILENCE provides 12.50 minutes of
silence.
CPR - Alarm will flash silent when CPR is engaged when unit is in use.
NOTE: Pressing ALARM SILENCE stops audible alarms; however, ALARM SILENCE LED will continue to flash until
alarm condition is cleared.
EXIT ALARM - An audible alarm will sound if a 10% or more decrease in patient weight is detected.
NOTE: On the Scale Control Panel, press EXIT ALARM button to turn Exit Alarm either ON or OFF.
5-2
HEATER
UP
TEMP
ON
°F
DOWN
HEAT
°C
OFF
ALARM
SILENCE