Air Adjust - Arjohuntleigh KinAir MedSurg Pulse User Manual

Table of Contents

Advertisement

Air Adjust

HEAD
20
°
18
°
16
°
14
°
12
°
10
°
8
°
6
°
4
°
2
°
INFLATE
HEAD
DEFLATE
AUTO PROGRAM AIR ADJUSTMENT
Use the Auto Program Air Adjust to automatically select air pressures for each section of cushions based on patient's weight
and height.
1.
Press AIR ADJUST button. The LED flashes when changes can be made.
2.
To make adjustments using the Auto Program preset air pressures:
Refer to the Auto Program Selection Guidelines below.
Locate the height value corresponding to that of the patient.
Look across table to column that relates to the patient's sex and note the weight corresponding with the patient's
height.
If the patient's weight is less than or equal to the value listed, press the TALL PATIENT button under the determined
patient weight range.
If the patient's weight exceeds the value listed, press the SHORT PATIENT button under the determined patient
weight range.
If the patient's weight falls on the borderline between weight ranges, it may be necessary to manually adjust the
established AUTO PROGRAM settings to suit the patient's requirements. For further instructions, refer to the
Manual Air Adjustment section in this chapter.
3.
Press the AIR ADJUST button to exit Air Adjust.
SHOULDER
BODY
LEG
20
20
20
°
°
°
18
18
18
°
°
°
16
16
16
°
°
°
14
14
14
°
°
°
12
12
12
°
°
°
10
10
10
°
°
°
8
8
8
°
°
°
6
6
6
°
°
°
4
4
4
°
°
°
2
2
2
°
°
°
INFLATE
INFLATE
INFLATE
SHOULDER
BODY
LEG
DEFLATE
DEFLATE
DEFLATE
AIR ADJUST
AUTO PROGRAM
PATIENT WEIGHT
GUIDELINES (LBS.)*
100
150
200
160
210
250
TALL
TALL
TALL
PATIENT
PATIENT
PATIENT
SHORT
SHORT
SHORT
PATIENT
PATIENT
PATIENT
5-4
AIR
ADJUST
HIGH
PULSE
FLASHING
LIGHT
INDICATES
NO PULSE
MODE
LOW
PULSE
*FOR ADDITIONAL HEIGHT
AND WEIGHT ADJUSTMENT
SEE REFERENCE GUIDE
ONE
BLOWER
AIR
ADJUST