Reset Totalizers; Reset Totalizers Using The Display - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

8.7

Reset totalizers

Display
ProLink III
Field Communicator Service Tools > Variables > Totalizer Control > Mass > Mass Total
Overview
When you reset a totalizer, the transmitter sets its value to 0. It does not matter whether
the totalizer is started or stopped. If the totalizer is started, it continues to track process
measurement.
Tip
When you reset a single totalizer, the values of other totalizers are not reset. Inventory values are not
reset.
8.7.1

Reset totalizers using the display

Prerequisites
The Totalizer Reset display function must be enabled.
The totalizer that you want to reset must be configured as a display variable. For example:
Procedure
Configuration and Use Manual
See
Section
8.7.1.
Device Tools > Totalizer Control > Totalizer and Inventories > Reset Mass Total
Device Tools > Totalizer Control > Totalizer and Inventories > Reset Volume Total
Device Tools > Totalizer Control > Totalizer and Inventories > Reset Gas Total
Device Tools > Totalizer Control > Totalizer and Inventories > Reset All Totals
Service Tools > Variables > Totalizer Control > Gas Standard Volume > Volume Total
Service Tools > Variables > Totalizer Control > Gas Standard Volume > GSV Total
Service Tools > Variables > Totalizer Control > All Totalizers > Reset All Totals
If you want to reset the mass totalizer, Mass Total must be configured as a display
variable.
If you want to reset the volume totalizer, Volume Total must be configured as a
display variable.
To reset the mass totalizer:
1. Scroll until the mass totalizer value appears.
2. Select.
Exit displays beneath the current totalizer value.
3. Scroll until Reset displays beneath the current totalizer value.
4. Select.
Transmitter operation
135