Transmitter operation
e.
Select again to confirm.
f.
Scroll to EXIT.
8.6
Reset mass and volume totalizers
Display
See Section 8.6.1 .
ProLink II
ProLink→Totalizer Control→Reset Mass Total
ProLink→Totalizer Control→Reset Volume Total
ProLink→Totalizer Control→Reset Gas Volume Total
ProLink→Totalizer Control→Reset
Field Communicator
Service Tools→Variables→Totalizer Control→Mass→Mass Total
Service Tools→Variables→Totalizer Control→Gas Standard Volume→Volume Total
Service Tools→Variables→Totalizer Control→Gas Standard Volume→GSV Total
Service Tools→Variables→Totalizer Control→All Totalizers→Reset All Totals
8.6.1
Reset mass and volume totalizers using the display
Prerequisites
To reset the totalizers using the display, this feature must be enabled and the appropriate process
variable (Mass Total, Volume Total, or Gas Volume Total) must be configured as a display variable.
Procedure
• To reset the mass totalizer using the display:
a.
Scroll until the mass totalizer value appears.
b.
Select.
c.
Scroll until RESET appears beneath the current totalizer value.
d.
Select.
e.
Select again to confirm.
f.
Scroll to EXIT.
g.
Select.
• To reset the volume (liquid or gas) totalizers using the display:
a.
Scroll until the volume totalizer value appears.
b.
Select.
c.
Scroll until RESET appears beneath the current totalizer value.
d.
Select.
e.
Select again to confirm.
f.
Scroll to EXIT.
g.
Select.
162
Micro Motion 9739 MVD Transmitters