5.2.2
Enable or disable Totalizer Reset from the display
Display
ProLink III
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Totalizer Reset
Overview
You can configure whether or not the operator is able to reset totalizers from the display.
Restrictions
•
•
•
Procedure
1.
2.
5.2.3
Enable or disable the Acknowledge All Alerts display
command
Display
ProLink III
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge
all alerts from the display.
Configuration and Use Manual
OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS RESET
Device Tools > Configuration > Totalizer Control Methods
This parameter does not apply to inventories. You cannot reset inventories from the display.
You cannot use the display to reset all totalizers as a group. You must reset totalizers
individually.
If the petroleum measurement application is installed, the operator must enter the off-line
password to perform this function, even if the off-line password is not enabled.
Ensure that the totalizers you want to reset have been configured as display
variables.
If the totalizer is not configured as a display variable, the operator will not be able to
reset it.
Enable or disable resetting the totalizer as desired.
Option
Description
Enabled
Operators can reset a totalizer from the display, if the totalizer is config-
ured as a display variable.
Disabled (default) Operators cannot reset totalizers from the display.
OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > ACK
Device Tools > Configuration > Transmitter Display > Ack All
Configure device options and preferences
73