When you reset an inventory, the transmitter sets its value to 0. It does not matter
whether the inventory is started or stopped. If the inventory is started, it continues to track
process measurement.
Tip
Mass and volume inventory totals cannot be set separately. They can only be reset together
simultaneously.
Prerequisites
To use ProLink III to reset the inventories, the feature must be enabled.
Once enabled, this feature remains enabled until it is disabled.
Configuration and Use Manual
To enable inventory reset in ProLink III:
1. Choose Tools > Options.
2. Select Reset Inventories from ProLink III.
3. Select OK.