Alert num-
ber
Alert title
A020
Calibration Factors
Missing
A021
Transmitter/Sensor/
Software Mismatch
A022
Configuration Data-
base Corrupt (Core
Processor)
A023
Internal Totals Cor-
rupt (Core Process-
or)
A024
Program Corrupt
(Core Processor)
A025
Boot Sector Fault
(Core Processor)
Configuration and Use Manual
Possible cause
Some calibration factors have not
been entered or are incorrect.
The configured board type does not
match the physical board, or the
configured sensor type does not
match the physical sensor.
Internal electronics failure.
Internal electronics failure.
Internal electronics failure.
Internal electronics failure.
Recommended actions
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
Verify the setting of the Sensor Type pa-
rameter.
If Sensor Type=Curved Tube, ensure that
no parameters specific to Straight Tube
have been set.
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
Ensure that the correct board is instal-
led.
Verify the setting of the Sensor Type pa-
rameter.
If Sensor Type=Curved Tube, ensure that
no parameters specific to Straight Tube
have been set.
Cycle power to the meter.
Contact customer support.
Cycle power to the meter.
Contact customer support.
Cycle power to the meter.
Contact customer support.
Cycle power to the meter.
Contact customer support.