Configuration and Use Manual
MMI-20016855
b) Select.
c) Scroll until START appears beneath the current totalizer value.
displays beneath the current totalizer value.
Exit
d) Select.
e) Select again to confirm.
f) Scroll to EXIT.
•
To stop all totalizers and inventories using the display:
a) Scroll until the word TOTAL appears in the lower left corner of the display.
Important
Because all totalizers are started or stopped together, it does not matter which total you use.
b) Select.
c) Scroll until STOP appears beneath the current totalizer value.
d) Select.
e) Select again to confirm.
f) Scroll to EXIT.
8.7 Reset totalizers
Display
See
ProLink III
Device Tools → Totalizer Control → Totalizer and Inventories → Reset Mass Total
Device Tools → Totalizer Control → Totalizer and Inventories → Reset Volume Total
Device Tools → Totalizer Control → Totalizer and Inventories → Reset Gas Total
Device Tools → Totalizer Control → Totalizer and Inventories → Reset All Totals
Field Communicator
Service Tools → Variables → Totalizer Control → Mass → Mass Total
Service Tools → Variables → Totalizer Control → Gas Standard Volume → Volume Total
Service Tools → Variables → Totalizer Control → Gas Standard Volume → GSV Total
Service Tools → Variables → Totalizer Control → All Totalizers → Reset All Totals
When you reset a totalizer, the transmitter sets its value to 0. It does not matter whether the totalizer is
started or stopped. If the totalizer is started, it continues to track process measurement.
Tip
When you reset a single totalizer, the values of other totalizers are not reset. Inventory values are not reset.
8.7.1 Reset totalizers using the display
Prerequisites
•
The Totalizer Reset display function must be enabled.
•
The totalizer that you want to reset must be configured as a display variable. For example:
Configuration and Use Manual
Reset totalizers using the display
Transmitter operation
June 2019
117