Measurement support
Tool
Field Communicator
If necessary, modify Zero Time. Zero Time controls the amount of time the transmitter
takes to determine its zero-flow reference point. The default Zero Time is 20 seconds. For
most applications, the default Zero Time is appropriate.
Postrequisites
Restore normal flow through the sensor by opening the valves. Verify that the sensor tubes
are full.
Need help?
•
•
•
•
9.7
Validate the meter
Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Flow
Overview
Meter validation compares flowmeter measurements reported by the transmitter to an
external measurement standard. If the transmitter value for mass flow, volume flow, or
density measurement is significantly different from the external measurement standard,
you may want to adjust the corresponding meter factor. The flowmeter's actual
measurement is multiplied by the meter factor, and the resulting value is reported and
used in further processing.
Prerequisites
Identify the meter factor(s) that you will calculate and set. You may set any combination of
the three meter factors: mass flow, volume flow, and density. Note that all three meter
factors are independent:
•
•
•
156
Path
Service Tools > Maintenance > Zero Calibration > Perform Auto Zero
If the zero fails:
Verify that there is no flow through the sensor, then retry.
Remove or reduce sources of electromechanical noise, then retry.
Set Zero Time to a lower value, then retry.
If the zero continues to fail, contact customer service.
OFF-LINE MAINT > CONFG > UNITS > MTR F
Device Tools > Configuration > Process Measurement > Flow
Device Tools > Configuration > Process Measurement > Density
Configure > Manual Setup > Measurements > Density
The meter factor for mass flow affects only the value reported for mass flow.
The meter factor for density affects only the value reported for density.
The meter factor for volume flow affects only the value reported for volume flow or
gas standard volume flow.
Micro Motion Model 2700 Transmitters with Intrinsically Safe Outputs