Programming Examples; Sample Transfers; Incremental Pipetting - Thermo Scientific Matrix MultiChannel Equalizer Pipette User Manual

With expandable, equal tip spacing
Table of Contents

Advertisement

C.3 Programming Examples
Examples of different pipetting sequences are described below, and are
ideal for use as training exercises. Press the keys shown in keycaps [ ],
then press the [Vol] key until the volume in parentheses ( ) is displayed. At
the end of the program, a triple beep tone
Recommendations for Accurate Pipetting for suggestions for better precision
and accuracy).
C.3.1 Sample Transfers
Example: Fill and dispense 100µl.
s
Vol
[Fill]
(100) [Enter]
t
s
Vol
[Disp]
(100) [Enter] [Enter]
t
The second option is a one-step program. Because the program ends at a
volume greater than zero, and the last step is not a mix, a "Purge" prompt
automatically appears at the end of the fill to dispense the liquid and return to
the beginning of the program.
C.3.2 Incremental Pipetting
Example: Fill and dispense 250 µl.
Option 1
s
Vol
[Fill]
(250) [Enter]
t
s
Vol
[Disp]
(10) [Enter]
t
s
Vol
[Disp]
(45) [Enter] [Enter]
t
Programming and Operation
is sounded. (see section D.3
Option 2
s
Vol
[Fill]
(100) [Enter] [Enter]
t
Note: The 10 µl Primer
Dispense reduces the
percentage of error in
incremental dispensing
(see section D.3
Recommendations for
Accurate Pipetting)
19

Advertisement

Table of Contents
loading

Table of Contents