Pipette Overview
Double beep tone: operation or programming key is not active in
current mode; insufficient dispense volume, purge required; illegal
programming step attempted (see Section E, Troubleshooting).
Triple beep tone: end of pipetting sequence; end of programming
sequence.
There are three options to choose from in setting the beep
tone. ON1 will have all tones sound at the appropriate times
listed above. ON2 will use only double and triple beep tones.
OFF will turn off all tones except for the tone to signify an illegal
programming step. Press the Special Function key twice to
display "TONE" and its current setting. To change the setting,
press the Volume key. If the beep tone is set for ON1 or ON2, the
beep tone symbol "
PACE
The pace controls the speed of repeat pipetting (paced
dispensing), FAST, SLOW or OFF. Press the Special Function key
three times to display "PACE" and its current setting. To change
the setting, press the Volume key.
Paced Dispense. When using the paced dispense feature, the
pipette dispenses the programmed volume at the specified pace
(fast or slow). Immediately after each dispense step, the pipette
emits a single beep tone (if beep tone is on) to prepare you for the
next dispense step (see Section C.5, Operating Procedures).
CAL
The calibration of the pipette is factory set for distilled water at
20-22 °C. If you are pipetting fluids with significantly different
specific gravities or temperatures, use this setting to recalibrate
the pipette. (See Section D, Calibration).
8
" appears at the bottom of the display.
Note: For paced dispensing, the beep tone should be set
ON1, to prepare you for each dispense step.