Thermo Scientific Matrix MultiChannel Equalizer Pipette User Manual page 49

With expandable, equal tip spacing
Table of Contents

Advertisement

Maintenance
3. Carefully Remove Console Cover.
Press and lift off console cover assembly (see Figure F.2e). BE CAREFUL not
to tear the attached flex cable! (Refer to Figure F.2f).
4. Unclasp Battery Connector. Lift one side of battery connector with a
small flathead screwdriver. Grasp all three wires and pull out connector
(see Fires F.2f,g).
CAUTION: Be careful not to tear the Flex Cable
Figure F.2f: Lift Battery Connector
5. Connect New Battery. Attach the wires of the new battery in place of the
old battery.
6. Reassemble Console Cover. Attach the pipette top to the main housing.
Fit the edges of the pipette top snugly within the edges along the top of the
pipette housing.
44
Figure F.2e: Remove Console Cover
Figure F.2g: Unclasp Battery Connector