4. Rotating Manifold Assembly After installing the piston/cylinder assembly,
install the manifold assembly into the main housing. Make sure manifold is
in closed position. To keep the tubing free from ejector contact, rotate the
manifold plate and tubing one complete turn counterclockwise (see Figure
F.1.4d). Spin tubing once, then reconnect tube one without twisting (it is the
only tube not twisted).
Note: The Manifold Assembly is comprised of the Manifold Plate and Tip
Fittings, and is connected to the Piston/Cylinder Assembly by the tubing.
5. Installing Manifold Assembly Place the back plane divider onto the
link assembly, then insert the manifold assembly into the main housing
support ribs. Be sure to inset the pin located on the last tip fitting of the
manifold assembly into the slot of the spacing indicator arm. Use a dowel
or screwdriver to move the tubing away from edges and points of contact
between the main housing and housing cover.
Figure F.1.4c: Installing Piston/Cylinder Assembly
Not twisted
Figure F.1.4d: Rotating Manifold Assembly
Maintenance