Recommended Equipment Decontamination Methods - Thermo Scientific Matrix MultiChannel Equalizer Pipette User Manual

With expandable, equal tip spacing
Table of Contents

Advertisement

G.2 Recommended Equipment Decontamination Methods
The following table lists the proper decontamination procedures for returning
Matrix Electronic pipettes to Thermo Fisher Scientific for service. Contact
customer service at 800-345-0206 (US) or +44(0) 161 486 2112 (Europe)
for a current equipment decontamination verification form.
Product/Component
Multichannel Pipettes
Pipette Exterior
Piston/Cylinder Assembly
Tip Manifold
Battery
Power Supply
Pipette Exterior
Single Channel Electronic Pipettes
Pipette w/Battery Removed
Lower cylinder
Piston (do not remove)
Power Supply
WARNING: Do not expose batteries to temperatures above 60 °C.
Heat at
65 °C/
70%/30%
10 Hr.
IPA/Water
YES
Wipe
YES
YES
YES
YES
NO
Wipe
NO
Wipe
NO
Wipe
YES
Wipe
YES
YES
YES
YES
NO
Wipe
Technical Service
10%
Ethylene
Bleach
Oxide
Wipe
YES
NO
YES
NO
YES
NO
NO
NO
NO
Wipe
NO
Wipe
YES
YES
YES
NO
YES
NO
NO
Autoclave
121 °C/
15 min.
NO
YES
YES
NO
NO
NO
NO
YES
NO
NO
47