Maintenance
4. Remove the ejector comb by sliding it down and forward off the tip fittings
(see Figure F.1.1b). Remove the plastic ejector trigger and spring. Next,
remove the back plane divider.
Figure F.1.1b: Removing Ejector Trigger and Ejector Comb
5. Remove the manifold assembly and piston/cylinder assembly by pulling the
motor screw down from the motor shaft coupling and out of the main housing.
Be careful not to misplace the motor screw washer. Remove the manifold
and piston/cylinder assemblies by firmly lifting them out of the main housing
support ribs (see Figure F.1.1c).
Figure F.1.1c: Removing Piston/Cylinder and Manifold Assemblies
33