Safety Precautions The Thermo Scientific™ mySPIN 6 Microcentrifuge is a personal mini-centrifuge designed for laboratories in the bioscience, medical, and chemistry fields. It may be used with biohazardous and/or hazardous materials. This instruction manual cannot address all safety hazards. It is the responsibility of the user to consult and observe all health and safety precautions and to assess the instrument’s suitability to the task.
Page 3
19. Close tube lids before starting centrifuge. Open lids can be ripped off during centrifugation and damage the unit. Open lids can cause fluid to be dispersed inside the centrifuge. 20. Do not insert non-standard tubes or other unfamiliar articles into the rotor. 21.
Page 4
WARNING The laboratory procedure should ensure that no person or hazardous substances are present within a 12” (30 cm) zone around the centrifuge when operating. WARNING If this product is not used as recommended by the manufacturer, the overall safety will be impaired.
Maximum RCF: 2,000 x g Electrical Power Adapter rating Input: 100-240 V AC 50/60 Hz, 0.8 A Output: 12 V DC mySPIN 6 Input: 12 V DC Current (under load): Approx. 0.6 A Power: Approx. 7.3 W Accessories: • Six tube adapters for 0.2 mL tubes and six tube adapters for 0.5 mL tubes; PN: 50146145 •...
EC Declaration of Conformity Model: Sprout ® Manufacturer: Heathrow Scientific Address: 620 Lakeview Parkway, Vernon Hills, IL 60061 USA European Contact: Emergo Europe Molenstraat 15, 2513 BH, The Hague, The Netherlands This unit has been constructed and conforms to the following: Safety Standards: IEC/EN 61010-1 IEC/EN 61010-2-020...
Turn power switch ON. If there is a smooth whirring sound and the unit accelerates with little or no vibration, mySPIN 6 is ready to use. If there are loud or unusual sounds, or excessive vibration, DO NOT OPERATE. Contact Thermo Fisher Scientific Technical Support.
If there is a smooth whirring sound and the unit accelerates with little or no vibration, mySPIN 6 is ready to use. If there are loud or unusual sounds, or excessive vibration, DO NOT OPERATE. Contact the Thermo Fisher Scientific Technical Support.
Maintenance mySPIN 6 is normally maintenance-free. Clean unit only when it is not plugged into an electrical outlet. When necessary, the housing and rotor can be wiped using a damp cloth and a mild, non-corrosive detergent. WARNING Do not use any solvent on the unit that may attack plastic or cause cracks in the rotor. Ensure the rotor is protected from corrosion and mechanical damage.