Calreset; Recommendations For Accurate Pipetting - Thermo Scientific Matrix MultiChannel Equalizer Pipette User Manual

With expandable, equal tip spacing
Table of Contents

Advertisement

D.2.5 CALRESET
CALRESET allows you to return to a default calibration step. To reset the
calibration, enter the CAL mode using the Special Function key. When the
calibration point is displayed, press the [Purge] key. The display flashes
"FACT," "CAL," then "SET" and the pipette resets to the default step. This
default setting is an average value for all pipettes of its class. This value may
differ from the factory-optimized calibration step documented on the
"Pipette Calibration Report" enclosed in this pipette's documentation.
D.3 Recommendations for Accurate Pipetting
Equalizer pipettes are factory calibrated to deliver accurate volumes with
distilled water at 20 °C. By using the calibration procedures described
above, you can pipette liquids of different specific gravity with a high degree
of accuracy and reproducibility. For optimal pipetting results, refer to the
following recommendations:
• Pipette at a constant temperature. Allow liquids and equipment
to equilibrate to ambient temperature. Volume aspiration with air
displacement pipettes varies with air pressure, relative humidity and the
liquids vapor pressure, all of which are temperature dependent.
• Attach tips with slow, even pressure. When applying tips, avoid
'banging' or applying excessive force to the tip(s), as this stretches the tip
seal and will affect volume. Use Matrix tips, which have been specifically
designed to match the tip fittings on Matrix pipettes.
• Calibrate the pipette for the liquid. If the liquid has a specific gravity
and vapor pressure equivalent to distilled water, calibrate using "EASY"
CAL mode. Otherwise, use "CALC" mode.
• Pre-wet the pipette tip. Aspirate and dispense a few times before
aspirating the liquid to increase humidity within the tip, reducing
sample evaporation that results in lower dispense volumes in initial few
pipettings. The need to pre-wet increases when working with liquids with
high vapor pressure and/or high surface tension.
• Immerse tips fully. Tips should be immersed approximately ¼ inch
(approx. 0.6 cm) below the liquid's meniscus and away from the
container surfaces to avoid aspiration of air and excess droplets clinging
to the outside of the tip.
• Discard a primer volume. When the motor changes from aspiration to
dispense, the percentage of error is greater in the first dispense than in
subsequent dispenses.
Calibration
27

Advertisement

Table of Contents
loading

Table of Contents