E. TROUBLESHOOTING
Possible procedural and system problems, as well as recommended solutions.
If you cannot resolve the problem, call technical service at 800-345-0206.
No power to the pipette.
Probable Cause
Battery is discharged.
Double beep tone
Probable Cause
An illegal programming step
is being entered, such as
programming two fills that
exceed the maximum fill
volume.
"ERR_1" is displayed. (Signifies one of the following conditions)
Probable Cause
Liquid is aspirated into the
pipette.
Motor is defective or the flex
circuit assembly is damaged.
Recommended Solution
Charge battery for at least one hour.
occurs.
Recommended Solution
Recheck the programming sequence.
Ensure you are not attempting to fill
above the maximum fill volume of
the pipette, or attempting to dispense
more than the total fill volume.
Recommended Solution
Disassemble and clean the pipette.
(See Section F.1 and/or F.2, for pipette
cleaning and autoclaving instructions).
Contact Thermo Fisher Scientific's
technical service at 800-345-0206

Troubleshooting

29