Matrix Equalizer Pipette Specifications - Thermo Scientific Matrix MultiChannel Equalizer Pipette User Manual

With expandable, equal tip spacing
Table of Contents

Advertisement

Pipette Overview
B.5 Equalizer Pipette Specifications
Item
Volume
No.
2032
250 µl
2034
1250 µl
2139
12.50 µl
2130
30 µl
2131
125.0 µl
2239
12.50 µl
2230
30 µl
2231
125 µl
Note: For Accuracy and Precision, values shown are expressed as a percent
(%) deviation or microliter (µl) value. When applied to desired volume, the
greater of the two values will always apply.
*Both values represent the deviation from the mean.
**Percentage values are expressed as the coefficient of variation.
Microliter values are expressed as the standard deviation.
10
Volume
Channel
Range
8-Ch
5-250 µl
8-Ch
15-1250 µl
8-Ch 384
0.50-12.50 µl
8-Ch 384
1.0-30.0 µl
8-Ch 384
2.0-125.0 µl
12-Ch 384
0.50-12.50 µl
12-Ch 384
1.0-30.0 µl
12-Ch 384
2.0-125.0 µl
Table B.5 Equalizer Pipette Specifications
Color
Accuracy*
Ring
(±)
Blue
2.5%/1.5 µl
Green
1.5%/6.0 µl
Red
2.5%/0.15 µl
Purple
2.0%/0.30 µl
Yellow
2.0%/1.0 µl
Red
2.5%/0.15 µl
Purple
2.0%/0.30 µl
Yellow
2.0%/1.0 µl
Precision**
(±)
0.7%/1.0 µl
0.6%/3.0 µl
2.0%/0.15 µl
1.7%/0.30 µl
1.0%/0.6 µl
2.0%/0.15 µl
1.7%/0.30 µl
1.2%/0.8 µl

Advertisement