2.5.8

Clearing the Memory

In SET-UP, press the MEM button:
The MEM symbol flashes
The small display indicates "ALL"
The main display indicates "cLR"
To clear the entire memory, press the MEM button again for 2 seconds:
The MEM symbol is displayed without flashing
The main display indicates "FrEE"
To clear the contents of a specific OBJ : TEST number:
Select the number using the
"cLr" remains displayed on the main display
Press the MEM button again for 2 seconds to clear the contents:
The small display indicates the OBJ : TEST number
The main display indicates "FrEE"
2.5.9

Communication Rate (RS-232)

In SET-UP, press the PRINT button.
The main display indicates the Communication Rate (300, 600, 1200,
2400, 4800, 9600 or Parallel).
"baud" appears on the small display. The value can be changed using
the
and
"Parallel" appearing in the display means that the parallel mode has
been selected. This is for printing on a parallel printer via the serial-to-
parallel adapter (RS-232 Centronics).

2.5.10 Lead Resistance Compensation

In SET-UP, press the
the small display. OFF can be selected with the
resistance will not be subtracted during continuity measurement.
To store the lead resistance:
Connect them together and press START (in SET-UP position
The resistance of the leads will be stored and indicated on the main display
NOTE:
This value is stored in memory, even when the instrument is switched OFF.
The lead compensation only comes into effect when performing continu-
ity measurements.
Megohmmeter Models 1050 /1060
and
buttons.
button three times. The
buttons
symbol and ON appears in
button. In this instance, the lead
)
19