AEMC 1227 User Manual

Thermo-anemometer data logger
Hide thumbs Also See for 1227:

Advertisement

Quick Links

1227
THERMO-ANEMOMETER
DATA LOGGER
ENGLISH
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for AEMC 1227

  • Page 1 ■ 1227 THERMO-ANEMOMETER DATA LOGGER ENGLISH User Manual...
  • Page 2 Statement of Compliance ® ® Chauvin Arnoux , Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards. We guarantee that at the time of shipping your instrument has met its published specifications.
  • Page 3: Table Of Contents

    CONTENTS PRECAUTIONS ..................................1 RECEIVING YOUR SHIPMENT ............................. 2 ORDERING INFORMATION ..............................2 1. GETTING STARTED ................................. 3 1.1. Battery Installation ..............................3 1.2. Instrument Components ............................4 1.3. Instrument Functions ............................... 4 1.4. Turning the Instrument ON/OFF ..........................5 1.5.
  • Page 4: Precautions

    Thank you for purchasing the Thermo-Anemometer Data Logger Model 1227. For best results from your instrument:  read these operating instructions carefully,  comply with the precautions for use. WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
  • Page 5: Receiving Your Shipment

    Adapter – U.S. wall plug to USB………….…..….…………………….………………..……………..Cat. #2153.78 Shock-proof housing …………………….….…………………………….….…….……….…………..Cat. #2122.31 Set of air flow measurement cones.……..……………….……………..………………………..…..Cat. #2122.32 Multifix Universal Mounting system………………….………………………...…………..……………..Cat. #5000.44 For the accessories and replacement parts, visit our web site: www.aemc.com Thermo-Anemometer Data Logger Model 1227...
  • Page 6: Getting Started

    1. Press the tab of the battery compartment cover and lift it clear. 2. Remove the battery compartment cover. 3. Insert the new batteries, ensuring correct polarity. 4. Close the battery compartment cover; ensuring it is completely and correctly closed. Thermo-Anemometer Data Logger Model 1227...
  • Page 7: Instrument Components

    10.Type B micro-USB connector 1.3. Instrument Functions The Model 1227 measures air speed from 0.25 to 35m/s, as well as volume flow from 0 to 99.999m3/h. It also measures temperature from 14 to 140°F (-10 to 60°C). This instrument is easy to use. It has extensive stand-alone capabilities and can: ...
  • Page 8: Turning The Instrument On/Off

    Fourth press displays minimum value. Fifth press returns to current values.  Long press exits MAX AVG MIN mode.  In MAP mode, pressing cycles through the maximum, average (mean), and minimum of the MAP measurements. Thermo-Anemometer Data Logger Model 1227...
  • Page 9: Display

    Recording, in MAX AVG MIN mode, in MAP mode, or in HOLD mode  Connected via the USB cable either to an external power supply or for communication with a computer  Communicating via Bluetooth  Set to Auto OFF disabled (see §2.4). Thermo-Anemometer Data Logger Model 1227...
  • Page 10: Setup

    7. When the instrument is connected, its name appears in the Data Logger Network in the left side of the Control Panel. A green check mark appears next to the instrument name indicating it is currently connected. Thermo-Anemometer Data Logger Model 1227...
  • Page 11: Instrument Date/Time

    When Auto OFF is disabled, the symbol appears in the instrument LCD screen. 2.5. Temperature Units button on the instrument front panel allows you to toggle between °C and °F as the temperature measurement unit. Thermo-Anemometer Data Logger Model 1227...
  • Page 12: Standalone Operation

    3.1.2. Air Flow (Volume) The Model 1227 can measure air flow both with and without using an optional air flow cone. Cones are typically used for measuring air flow through grated ventilation registers and other passages where the flow is disrupted and distorted, resulting in inconsistent measurements.
  • Page 13 7. The air flow and temperature measurements are displayed on the LCD. (Wait for the measurements to stabilize before accepting the reading.) Always hold the sensor by its handle. Keep your fingers away from the sensor propeller. Thermo-Anemometer Data Logger Model 1227...
  • Page 14: Hold Function

    To enter MAP mode, press for >2 seconds. Place the sensor at the first measurement point and press record the value in memory. The MAP counter is incremented. Repeat this step for all the other points of the map. Thermo-Anemometer Data Logger Model 1227...
  • Page 15: Recording Measurements

    The firmware version is incompatible with the instrument. Reload the previous firmware version. Er.13 Recording scheduling error. Ensure that the instrument’s time and the time of the DataView Data Logger Control Panel are the same (see §2.3). Thermo-Anemometer Data Logger Model 1227...
  • Page 16: Dataview

    Download recorded data from the instrument to the computer.  Generate reports from downloaded data.  View instrument measurements in real time on the computer. For information about performing these tasks, consult the DataView Data Logger Control Panel Help. Thermo-Anemometer Data Logger Model 1227...
  • Page 17: Technical Characteristics

    Intrinsic uncertainty ± 8% R ± 8% R Specified measurement range 0.00 to 2.99 cfm 3.0 to 99.9 cfm 100 to 2999 cfm Resolution 0.01 cfm 0.1 cfm 1 cfm Intrinsic uncertainty ± 8% R Thermo-Anemometer Data Logger Model 1227...
  • Page 18: Memory

    Operating range: 14 to 140°F (-10 to 60°C) and 10 to 90%RH without condensation Storage range: -4 to 158°F (-20 to 70°C) and 10 to 95%RH without condensation, without batteries Altitude: < 6562’ (< 2000m), and 32,800’ (10,000m) in storage. Pollution degree: 2 Thermo-Anemometer Data Logger Model 1227...
  • Page 19: Mechanical Characteristics

    Inrush protection: IP 50, with the USB connector closed, per IEC 60 529. Drop impact test 3.28’ (1m) per IEC 61010-1. 5.9. Compliance with International Standards The instrument is compliant with standard IEC 61010-1. 5.10. Electromagnetic Compatibility (CEM) The instrument is compliant with standard IEC 61326-1. Thermo-Anemometer Data Logger Model 1227...
  • Page 20: Maintenance

    Spent batteries must not be treated as ordinary household waste. Take them to an appropriate recycling collection facility. 6.3. Firmware Update AEMC may periodically update the instrument’s firmware. Updates are available for free download. To check for updates: 1. Connect the instrument to the Data Logger Control Panel.
  • Page 21: Repair And Calibration

    ® ® Contact: Chauvin Arnoux , Inc. d.b.a. AEMC Instruments Phone: (800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) Fax: (603) 742-2346 E-mail: techsupport@aemc.com Thermo-Anemometer Data Logger Model 1227...
  • Page 22: Limited Warranty

    LIMITED WARRANTY Your AEMC instrument is warranted to the owner for a period of two years from the date of original purchase ® against defects in manufacture. This limited warranty is given by AEMC Instruments, not by the distributor from whom it was purchased.
  • Page 23 NOTES: Thermo-Anemometer Data Logger Model 1227...
  • Page 24 01/18 99-MAN 100449 v1 ® ® Chauvin Arnoux , Inc. d.b.a. AEMC Instruments 15 Faraday Drive • Dover, NH 03820 USA • Phone: (603) 749-6434 • Fax: (603) 742-2346 www.aemc.com...