2.6.8 Clearing the Memory

In SET-UP, press the MEM button:
■ The MEM symbol flashes.
■ The small display indicates "ALL."
■ The main display indicates "cLR."
To clear the entire memory, press the MEM button again for 2 seconds:
■ The MEM symbol is displayed without flashing.
■ The small display indicates "ALL."
■ The main display indicates "FrEE."
To clear the contents of a specific OBJ : TEST number:
■ Select the number using the
■ "FrEE" or "OCC" appears on the main display.
Press the MEM button again for 2 seconds to clear the contents:
■ The small display indicates the OBJ : TEST number.
■ The main display indicates "FrEE."

2.6.9 Communication Rate (USB Port)

■ In SET-UP, press the PRINT button.
■ The main display indicates the Communication Rate (e.g. 300, 600, 1200,
2400, 4800, 9600, or Parallel).
■ "bAUd" appears on the small display. The value can be changed using the ▲
and ▼ buttons. For the USB link to operate, the baud rate must be 9600.
■ "Parallel" is not available for use.

2.6.10 Default Device Configuration

In SET-UP, press the
■ "SEt" appears in the small display.
■ "DFLt" (flashing) appears in the main display.
Press START to select the default configuration settings (see the previous table).

2.6.11 Voltage Disturbance Limit

In SET-UP, press the
■ "dISt" appears in the small display.
■ "003U" (flashing) appears in the main display.
■ Modify this value if necessary using the
0.10 - 0.20 - 0.03).
Megohmmeter Models 5050 & 5060
and
buttons.
button three times:
button five times:
button (possible choice between:
23