Power Supply - AEMC 1060 User Manual

Megohmmeter
Hide thumbs Also See for 1060:
Table of Contents

Advertisement

Range
0.01 to 39.99Ω
40.0 to 399.9Ω
0.400 to 3.999kΩ
4.00 to 39.99kΩ
40.0 to 399.9kΩ

3.6 Power Supply

8 x 1.5V (C cell) alkaline batteries; LR14 (Model 1050)
Rechargeable battery NiMH (Model 1060)
Recharge: 85 to 256V / 50-60Hz (electrical safety: 256V, Cat. III)
Model 1050:
Measurement: Average operating time
Insulation Measurement:
21,000 x 5s measurements with 20s pause, for normal load (1000V/1mA)*
Continuity Measurement:
16,000 measurements lasting 5s with pause of 20s at nominal load
(1000V/1mA)*
*Assuming that a 10 minute long PI measurement is made 5 times a day, the operating time
will be 67 days (10 weeks or 2.5 months).
Model 1060:
Measurement: Average operating time
Insulation Measurement:
5000 x 5s measurements with 29s pause, for normal load (1000V/1mA)*
Continuity Measurement:
4000 measurements lasting 5s with pause of 20s at nominal load (1000V/1mA)*
*Assuming that a 10 minute long PI measurement is made 5 times a day, the operating time
will be 16 days (2.5 weeks or 0.5 months).
Recharging Time (1060)
4.5 hours to reach 100% of its capacity (max recharge time: 6 hours)
0.5 hours to reach 10% of its capacity (operating radius: 1 day approx)
Megohmmeter Models 1050 /1060
Resolution
0.01Ω
0.1Ω
10Ω
100Ω
Accuracy
±3% R ± 3cts
29