Run Temperature; Avanti J Series Centrifuges; Microprocessor-Controlled J6 Series Centrifuges; Relative Centrifugal Fields - Beckman Coulter JS-5.2 Instructions For Use Manual

Swinging-bucket rotors for use in j6 series/avanti jxn-26, j-26xp series centrifuges
Table of Contents

Advertisement

JS-5.2 and JS-4.0 Swinging-Bucket Rotors
Preparation and Use
Table 5 Relative Centrifugal Fields
Rotor Speed
(RPM)
5200
5000
4500
4000
3500
3000
2600
2500
2000
1500
1000
a. Entries in this table are calculated from the formula RCF = 1.12 r (RPM/1000)2 and then rounded to three significant digits.

Run Temperature

To ensure that the rotor reaches the required temperature during the run, follow the appropriate
instructions below for the centrifuge model being used.

Avanti J Series Centrifuges

Enter the run temperature according to the instructions in your centrifuge instruction manual.
No additional input is required.

Microprocessor-Controlled J6 Series Centrifuges

Enter rotor code
1
Press the
temperature.
2
Find the compensation value in
run speed.
• The compensation values listed in
centrifuge models.
1-22
a
Relative Centrifugal Field ( g)
Buckets
At r
At r
max
av
(226 mm)
(156 mm)
6840
4720
6330
4370
5130
3540
4050
2800
3100
2140
2280
1570
1710
1180
1580
1090
1010
698
569
393
253
174
5.2
(for the JS-5.2) or rotor code
TEMP
key on the centrifuge control panel and then use the keypad to enter the sample
Table 6
Microplates
At r
At r
min
max
(86 mm)
(192 mm)
2600
2410
1950
1540
1180
866
651
1459
602
1340
385
860
216
483
96
215
4.0
(for the JS-4.0) and then follow the steps below.
that corresponds with the set temperature and
Table 6
are approximates for all J6 series
At ra
At r
v
min
(178 mm)
(164 mm)
1350
1240
1250
1150
797
734
448
413
199
183
J6-TB-006AP

Advertisement

Table of Contents
loading

This manual is also suitable for:

Js-4.0

Table of Contents