Switching the unit ON and OFF

NOTE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
Within the machine, grease depots are included for lubricating
components that can cause error messages at low temperatures.
ü Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry (for at least one hour)
Ä The unit is dry and can be put into operation.
NOTE
Do not adjust the line voltage
The unit automatically adjusts to the line voltage.
Switch on the unit
ü The milling and grinding unit is connected to the power supply.
1. The main switch on the rear side of the unit is set to position I (ON).
2. Press the ON/OFF button on the front panel.
Ä The unit switches on and the display lights up.
Power-up display
When the milling and grinding unit is switched on, the display shows a
picture of the milling and grinding unit trying to contact the PC.
You can start or stop the water pump by pressing the "Pump" button
(C). This enables you to drain the water circuit without connecting to the
PC (e.g. prior to transport) or fill the water circuit during startup.
You can call up the IP address by pressing the "Config" button (D). You
can configure the milling and grinding unit in the network with this
address.
Switching the unit off
ü The unit has finished the machining operation.