Table of Contents Dentsply Sirona Operating Instructions CEREC Primescan AC, Primescan AC Table of Contents General data ......................Dear Customer,.................... Contact information..................General information about these operating instructions ......General conventions and structure of the document ........1.4.1 Structure of the document..............
Page 3
Dentsply Sirona Table of Contents Operating Instructions CEREC Primescan AC, Primescan AC Controls and functional elements..............3.3.1 Overview of the front panel ............3.3.2 Overview of the rear panel............. Operating state LED ..................Certification ....................Electromagnetic compatibility ..............3.6.1 Electromagnetic emission ..............
Page 4
Table of Contents Dentsply Sirona Operating Instructions CEREC Primescan AC, Primescan AC Maintenance ......................Cleaning, disinfection, and sterilization............6.1.1 Cleaning agents and disinfectants ..........6.1.1.1 Cleaning agents............6.1.1.2 Wipe disinfectants (virucidal limited) ......6.1.1.3 High-level disinfectant ..........6.1.2 Uncritical surfaces including monitor ..........
Operating Instructions CEREC Primescan AC, Primescan AC 1.1 Dear Customer, General data Dear Customer, Thank you for your purchase of this CEREC Primescan AC / Primescan ® unit from Dentsply Sirona. The CEREC Primescan AC enables you to produce dental restorations, e.g.
1 General data Dentsply Sirona 1.3 General information about these operating instructions Operating Instructions CEREC Primescan AC, Primescan AC General information about these operating instructions Observe the Operating Instructions Please familiarize yourself with the unit by reading through these Operating Instructions before putting it into operation. It is essential that you comply with the specified warning and safety information.
Dentsply Sirona 1 General data Operating Instructions CEREC Primescan AC, Primescan AC 1.4 General conventions and structure of the document General conventions and structure of the document 1.4.1 Structure of the document 1.4.1.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these instructions for use.
1 General data Dentsply Sirona 1.4 General conventions and structure of the document Operating Instructions CEREC Primescan AC, Primescan AC 1.4.2 Operating conventions Example Meaning Tapping Pressing once and releasing the finger or the left trackball/touchpad key on the acquisition unit.
Exclusion of liability If the operator does not meet the obligation to carry out such maintenance or fault messages are ignored, Dentsply Sirona or its authorized dealer does not assume any liability for damage caused. 66 54 573 D3696...
Operating Instructions CEREC Primescan AC, Primescan AC Intended use The CEREC Primescan AC / Primescan AC acquisition unit creates digital impressions, which are sent to a laboratory and/or in connection with the milling unit are used to manufacture computer-supported dental restorations, e.g.
Storage battery pack recycling symbol (see "Disposal of the storage battery pack [→ 88]") The CEREC Primescan AC / Primescan AC acquisition unit may con- tain an RF transmitter in the form of a WLAN card or a separate wire- less module.
1 General data Dentsply Sirona 1.7 Legend Operating Instructions CEREC Primescan AC, Primescan AC Symbols on the packaging Take note of the following symbols on the packaging: Protect from moisture Fragile; handle with care Temperature during storage and transport Relative humidity during storage and transport...
In extreme cases, it will be impossible to establish a wireless connection at all. Dentsply Sirona has selected the best possible configuration for data communication via the wireless interface (CEREC Radio Device) or WLAN, which generally ensures perfect functioning of this connection.
CAUTION Note on the prevention, recognition, and elimination of unintended electromagnetic effects: The CEREC Primescan AC / Primescan AC acquisition unit is Class B equipment (classified according to CISPR 11, EN 60601-1-2: 2015 based on IEC 60601-1-2: 2014). This unit may be used in professional equipment of health services.
Dentsply Sirona 2 Safety instructions Operating Instructions CEREC Primescan AC, Primescan AC 2.2 Basic safety information NOTE Danger posed by broken glass Stress to the glass surfaces of the control console and monitor from strong forces and impacts must be prevented, otherwise there is a risk of the glass breaking.
2 Safety instructions Dentsply Sirona 2.2 Basic safety information Operating Instructions CEREC Primescan AC, Primescan AC 2.2.4 Movement and stability of the unit NOTE The unit can overturn or slip away For reasons of tilt stability, the unit must be pulled by its front or rear handle when being moved.
In order to ensure product safety, this device may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. In particular, only the power cable also supplied or the corresponding original spare part may be used with the unit.
2 Safety instructions Dentsply Sirona 2.3 Safety labels Operating Instructions CEREC Primescan AC, Primescan AC Plug connections of external interfaces CAUTION Adaptation of acquisition unit to external components Additional devices connected to external interfaces must be tested according to the relevant standards, e.g.: EN 60601-1:2006 + Cor.:2010 + A1:2013,...
Dentsply Sirona 2 Safety instructions Operating Instructions CEREC Primescan AC, Primescan AC 2.4 Electrostatic charge Electrostatic charge 2.4.1 ESD warning labels ESD warning label CAUTION Risk of injury or damage to components from electrostatic discharge For electrical components labeled with an ESD warning label, observe the following instructions.
2 Safety instructions Dentsply Sirona 2.5 Wireless phone interference with equipment Operating Instructions CEREC Primescan AC, Primescan AC Formation of an electrostatic charge Electrostatic charges generally build up whenever two bodies are rubbed against each other, e.g. when walking (shoe soles against the floor) or driving a vehicle (tires against the street pavement).
Dentsply Sirona 2 Safety instructions Operating Instructions CEREC Primescan AC, Primescan AC 2.6 Integration in a network or connection to a modem Integration in a network or connection to a modem NOTE Observe the following installation regulations The following installation regulations apply to integration of the...
2 Safety instructions Dentsply Sirona 2.7 Disposal Operating Instructions CEREC Primescan AC, Primescan AC Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
3 Product description Dentsply Sirona 3.1 Technical description Operating Instructions CEREC Primescan AC, Primescan AC PC hardware Special PC with the following equipment: ® ● Processor: Intel ● RAM: 32GB RAM ● Hard disks: 1x PCIe SSD, 1x SATA HDD ●...
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.2 Technical data Technical data Type designation CEREC Primescan AC / Primescan AC Rated line voltage 100 - 240 VAC /50 - 60 Nominal current 5.0 – 2.1 A...
3 Product description Dentsply Sirona 3.2 Technical data Operating Instructions CEREC Primescan AC, Primescan AC Dimensions and weight Dimensions W x H x D in mm 408 (537) x 1190 x 443 in inches 16.06 (21.14) x 46.85 x 17.44 Weight ●...
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.3 Controls and functional elements Controls and functional elements 3.3.1 Overview of the front panel NOTE CEREC Primescan / Primescan is calibrated The CEREC Primescan / Primescan scanner is calibrated ex works.
3 Product description Dentsply Sirona 3.3 Controls and functional elements Operating Instructions CEREC Primescan AC, Primescan AC Acquisition unit with trackball CEREC Primescan / Center trackball button Primescan 4 castors with parking Left trackball button brake Foot switch/foot pedal Trackball...
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.3 Controls and functional elements 3.3.2 Overview of the rear panel Fuses Power cable holder Main switch On button (operating state I = ON, 0 = OFF LED integrated)
3 Product description Dentsply Sirona 3.4 Operating state LED Operating Instructions CEREC Primescan AC, Primescan AC Operating state LED To position the operating state LED, see chapter “Overview of the rear panel [→ 29]”. LED lights up blue: Acquisition unit is connected with the socket.
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.5 Certification Certification CE mark This product bears the CE mark in accordance with the provisions of the Council Directive 93/42/EEC of June 14, 1993 concerning medical devices and their changes.
EMC aspects. CEREC Primescan AC / Primescan AC complies with the requirements for electromagnetic compatibility (EMC) according to EN 60601-1-2: 2015 based on IEC 60601-1-2: 2014. CEREC Primescan AC / Primescan AC is hereinafter referred to as "UNIT". 3.6.1 Electromagnetic emission The UNIT is intended for operation in the electromagnetic environment specified below.
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.6 Electromagnetic compatibility 3.6.2 Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
Dentsply Sirona 3 Product description Operating Instructions CEREC Primescan AC, Primescan AC 3.6 Electromagnetic compatibility Immunity to interference against high-frequency electromagnetic fields in the direct vicinity of wireless communi- cation devices IEC 61000-4-3 Test frequency Modulation Required immunity test Maintained immunity test level (V/m)
3 Product description Dentsply Sirona 3.6 Electromagnetic compatibility Operating Instructions CEREC Primescan AC, Primescan AC 3.6.3 Working clearances Recommended working clearances The UNIT is intended for operation in an electromagnetic environment, between portable and mobile RF where radiated RF interference is checked. The customer or the user of...
4.1 Transport and unpacking Installation and startup Transport and unpacking All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery. 1. Check the delivery note to ensure that the consignment is complete.
➢ Always place the sensitive scanner in its holder! NOTE Risk of damage posed by pulling on the scanner cable If the CEREC Primescan AC / Primescan AC is moved by pulling on the scanner cable, there is a risk of damage to the cable, scanner and device.
Dentsply Sirona 4 Installation and startup Operating Instructions CEREC Primescan AC, Primescan AC 4.4 Initial startup Notes on network installation The network card is installed. The cable with the RJ-45 connectors establishes the network connection or is connected to the CEREC radio module.
4 Installation and startup Dentsply Sirona 4.4 Initial startup Operating Instructions CEREC Primescan AC, Primescan AC 4.4.3 Insert battery (optional) CAUTION Risk of fire or chemical injury Improper handling of the storage battery used in this unit can lead to a risk of fire or chemical injury.
Dentsply Sirona 4 Installation and startup Operating Instructions CEREC Primescan AC, Primescan AC 4.4 Initial startup 2. Slide in the storage battery using the guide rails and guide pins up to the stop and screw it in place with 2 fastening screws (C).
CAUTION Use only the supplied power cord Use only the power cord supplied by Dentsply Sirona to connect the acquisition unit to the power supply. If the acquisition unit is switched on at the main switch, then it can be switched on at the ON button.
Dentsply Sirona 4 Installation and startup Operating Instructions CEREC Primescan AC, Primescan AC 4.4 Initial startup 3. Switch the milling unit on (see the Operating Instructions for the milling unit). 4. After loading the operating system, start the CEREC SW / Connect SW application by double-clicking on the CEREC SW / Connect SW icon.
4 Installation and startup Dentsply Sirona 4.4 Initial startup Operating Instructions CEREC Primescan AC, Primescan AC 4.4.6 Battery-backed operation (optional) Introduction The acquisition unit PC has a battery-backed power supply. It is thus possible to operate the acquisition unit for a short time with no line voltage connected.
At least every six weeks the CEREC Primescan AC / Primescan AC acquisition unit should be connected to the mains supply to charge the storage batteries.
4 Installation and startup Dentsply Sirona 4.4 Initial startup Operating Instructions CEREC Primescan AC, Primescan AC Monitoring program The monitoring program is represented in the task bar by a symbol: The color of the symbol has the following meaning: ● Blue: –...
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.1 Working with the touch monitor Operation Working with the touch monitor 5.1.1 Adjusting the position of the monitor You can guide the touch monitor to a position convenient for you. In doing so, there is the option to adjust the height/tilt position and a lateral pivot.
5 Operation Dentsply Sirona 5.1 Working with the touch monitor Operating Instructions CEREC Primescan AC, Primescan AC 5.1.2 Touch functionality You can activate the touch function with and without gloves. The following gestures are supported: Edit a 3D model with multi-touch You can edit the 3D model using multi-touch.
5.1 Working with the touch monitor 5.1.3 Adjusting touch-sensitivity and buzzer volume The CEREC Primescan AC / Primescan AC provides you with the following options: ● Adjusting the touch-sensitivity of the touchpad/trackball buttons ● Adjusting the volume for their acoustic feedback 1.
5 Operation Dentsply Sirona 5.2 Optical impressions with the scanner Operating Instructions CEREC Primescan AC, Primescan AC Optical impressions with the scanner CAUTION Hot tip of the scanner sleeve! The tip of the scanner sleeve is continuously heated up! The surface temperature of the sleeve may be as high as 51°C.
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.2 Optical impressions with the scanner CAUTION Potentially hazardous optical radiation The scanner transmits potentially hazardous optical radiation which may cause harm to the eyes. ➢ During operation, do not look directly at the scanner for long periods.
5 Operation Dentsply Sirona 5.2 Optical impressions with the scanner Operating Instructions CEREC Primescan AC, Primescan AC Proceeding with scanning procedure 1. Activate the foot control or click on the switch with the cursor. Ä The scanning procedure begins. 2. Proceed with the scanning procedure as described above.
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.3 Scanner guide Scanner guide CAUTION After each use Prepare the scanner once again after each patient. ➢ Follow the instructions in the "Scanner [→ 63]" section on cleaning, disinfection and sterilization in order to avoid cross- contamination between patients.
5 Operation Dentsply Sirona 5.3 Scanner guide Operating Instructions CEREC Primescan AC, Primescan AC 5.3.1 Occlusal scan Important: Ensure that the distance between the coated sapphire glass of the scanner and the scanned surface is observed. The distance must be between 0 - 2 0 mm (ideally: 2 m m). The scanner does not rest on the teeth or the gums.
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.3 Scanner guide 5.3.3 Lingual scan ü The scanner is on the tooth that is positioned next to the preparation in the distal direction. 1. Rotate the scanner to maximum 20° toward the lingual direction.
5 Operation Dentsply Sirona 5.3 Scanner guide Operating Instructions CEREC Primescan AC, Primescan AC 5.3.5 Single and multiple buccal registration The buccal registration establishes the allocation of jaw exposures. ü The jaw with the preparation is scanned. 1. Scan the occlusal, buccal and lingual view of the antagonist (see the section “Occlusal scan ...
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.3 Scanner guide 5.3.6 Square and full jaw scan You can use different scanning procedures for scanning a quadrant or a full jaw. Find two procedures as follows to help you gain access should such help be necessary.
5 Operation Dentsply Sirona 5.3 Scanner guide Operating Instructions CEREC Primescan AC, Primescan AC Procedure 2 1. Start occlusally on the distal tooth, tilt the scanner approx. 60° in an oral direction and move it orally along the dental arch up to the opposite distal tooth.
Dentsply Sirona 5 Operation Operating Instructions CEREC Primescan AC, Primescan AC 5.4 Software for the scanner Software for the scanner Only use software which is pre-installed on the system. 5.4.1 Cut out model areas With the "Cut" function, you can should be able to cut out model areas.
Some countries have legal regulations which require regular safety inspections of electrical devices or systems by the operator. Dentsply Sirona would like to draw your attention to the fact that a so- called "retest" (repeat test) must be carried out for the CEREC Primescan AC / Primescan AC acquisition unit at least every three years.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization Cleaning, disinfection, and sterilization NOTE Approved cleaning agents and disinfectants Use only cleaning and disinfecting agents which have been approved by Dentsply Sirona! 6.1.1 Cleaning agents and disinfectants 6.1.1.1...
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 6.1.2 Uncritical surfaces including monitor NOTE Do not allow liquids to penetrate into the ventilation slots! NOTE Never use corrosive cleaning agents, wax or solvents.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6.1.3 Trackball holder (if available) NOTE No residues in the spherical calotte The spherical calotte is designed without openings and may therefore come into contact with fluids.
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 6.1.4.1 General information The CEREC Primescan / Primescan scanner is a very sensitive optical device and must therefore be handled with the utmost care. Protect the...
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6.1.4.2 Components of the scanner CAUTION Risk of cross-contamination Without the window sleeve / disposable sleeve the scanner must not be used in the patient’s mouth. Without the window sleeve / disposable sleeve, the scanner cannot be disinfected or sterilized and can lead to cross-contamination.
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 6.1.4.3 Removing the sleeve If it is necessary to remove the sleeve, proceed as follows: 1. Grip the sleeve in the marked area. NOTE There is a risk of damaging the scanner window or the sleeve window if the sleeve is not pushed straight toward the front.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6.1.4.6 Wipe disinfection of the scanner and window sleeve NOTE Do not spray the CEREC Primescan / Primescan scanner or immerse it in cleaning agents or disinfectants! 1.
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 6.1.4.7 High-level disinfection of the window sleeve NOTE The hot air sterilization and high-level disinfection must not be combined. The complete process for high-level disinfection (HLD) is as follows –...
Page 69
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6. Push the plug completely into the metal bracket. 7. Press the sleeve onto the plug while holding the metal bracket firmly in place in order to ensure that no particle contamination or fluids penetrate the inside of the sleeve.
Page 70
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 9. Insert the metal bracket with the sleeve. 10. Place the cover on the container and leave the sleeve in the ® disinfectant for at least 12 minutes (CIDEX OPA).
Page 71
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 11. Remove the cover and take the bracket with the sleeve out of the container. 12. Carefully rinse the sleeve with tap water for at least 30 seconds while holding it with the bracket.
Page 72
Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 17. Dry the entire outer surface of the sleeve as well as the inner section around the plug using a soft lint-free cloth. Ensure that no fluid is able to penetrate the inside of the sleeve.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6.1.4.8 Hot air sterilization of the window sleeve NOTE Hot air sterilization and high-level disinfection must not be combined. The process for hot air sterilization is as follows: 1.
6 Maintenance Dentsply Sirona 6.1 Cleaning, disinfection, and sterilization Operating Instructions CEREC Primescan AC, Primescan AC 6.1.4.9 Using disposable sleeves The CEREC Primescan / Primescan scanner can be operated with disposable plastic sleeves in order to ensure maximum infection control.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.1 Cleaning, disinfection, and sterilization 6.1.5 Cleaning the inside of the sleeve window If you should realize during the scanning process that the inside of the sleeve is dirty, then follow the following process: 1.
6 Maintenance Dentsply Sirona 6.2 Calibrating the scanner Operating Instructions CEREC Primescan AC, Primescan AC Calibrating the scanner The measurement procedure used by the system requires the use of a calibrated scanner. The scanner is calibrated ex works. Calibrate the scanner after every reinstallation and after each time that it is transported.
Page 77
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.2 Calibrating the scanner Calibrating the scanner 1. Remove the protective cap from the calibration set. 2. Mount the calibration set on the tip of the scanner until it locks into place.
Page 78
6 Maintenance Dentsply Sirona 6.2 Calibrating the scanner Operating Instructions CEREC Primescan AC, Primescan AC Measuring the position of the exit window 1. Mount the bottom side of the calibration set to the tip of the scanner. 2. Click on the "OK" button.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.3 Color calibration Color calibration General information NOTE Faulty color analysis The color analysis can be negatively impacted due to strong light incidence and it can lead to varying results.
Page 80
6 Maintenance Dentsply Sirona 6.3 Color calibration Operating Instructions CEREC Primescan AC, Primescan AC Color-calibrating the scanner NOTE Only use color calibration set with clean, dry CEREC Primescan / Primescan scanner In order to achieve optimum results, the CEREC Primescan / Primescan scanner must be clean, disinfected and dry before color calibration.
Page 81
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.3 Color calibration Replacing the color calibration set NOTE Regularly replacing the color calibration set In order to achieve optimum results, the color calibration set must be replaced regularly.
6 Maintenance Dentsply Sirona 6.4 Replacing the main fuse Operating Instructions CEREC Primescan AC, Primescan AC Replacing the main fuse DANGER Potentially lethal shock hazard People can be injured or electrical components of the unit destroyed. ➢ Switch off the unit prior to beginning work.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.5 Charge battery (optional) Charge battery (optional) NOTE Information on back-up cycles The storage battery is designed for fully cable-free use of Scan, Design and Grinding at a power requirement of 250 W for 60 minutes.
6 Maintenance Dentsply Sirona 6.6 Replace battery (optional) Operating Instructions CEREC Primescan AC, Primescan AC Replace battery (optional) Service cover 2 guide rails Storage battery 2 guide pins 2 fastening screws 1. Open the service cover on the back panel.
Dentsply Sirona 6 Maintenance Operating Instructions CEREC Primescan AC, Primescan AC 6.7 Replacing the O-ring Replacing the O-ring Removing the worn O-ring ü The sleeve has been removed from the scanner body. 1. Hold the holding ring between your thumb and index finger.
Page 86
6 Maintenance Dentsply Sirona 6.7 Replacing the O-ring Operating Instructions CEREC Primescan AC, Primescan AC 1. Slide the O-ring at one point into the O-ring groove. 2. Starting from this point, carefully use your thumb and index finger to push the O-ring into the groove (in a similar way to pushing it out but without any force).
Dentsply Sirona 7 Disposal Operating Instructions CEREC Primescan AC, Primescan AC Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
7 Disposal Dentsply Sirona 7.1 Disposal of the storage battery pack Operating Instructions CEREC Primescan AC, Primescan AC Disposal of the storage battery pack The storage battery pack must be subjected to recycling if it becomes defective or reaches the end of its service life. For country-specific information on disposal, contact your local dental dealers.
Dentsply Sirona 8 Appendix Operating Instructions CEREC Primescan AC, Primescan AC 8.1 Making backup copies Appendix Making backup copies To increase the system's data security and protect themselves against data losses, users should make backup copies of the data regularly.