5 Installation and startup
5.4 Commissioning
32
5.4.8.2

Water tank CEREC MC XL Basic

A
Water tank
A
Filter insert
B
Tank cap
ü The water tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off.
NOTE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter insert is completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait for a short time until the filter insert is completely soaked; then
add an appropriate amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see Switching the unit ON and OFF  [→ 33]).
9. Switch the pump on (press the "Pump" button) to fill the water
circuit.
10. Fill the water tank up again until the filter insert is completely
immersed (up to the bottom edge of the cap thread).