Operation

5.1

Acquiring the X-ray image

5.1.1

Switching the unit on and starting the software

5.1.1.1

Switching the unit on

CAUTION
Malfunctions can occur when the unit is switched on.
A patient positioned in the unit may be injured by moving parts.
➢ Please ensure that a patient is not positioned in the unit when
activating the unit and selecting the operating mode (up until the
completion of sensor positioning).
NOTE
Fluctuations in temperature can cause condensation to form in the unit.
Electrical components are destroyed by short circuits.
➢ Do not switch the unit on until the temperature of the unit has
adapted to the ambient temperature and the condensation has
evaporated. See the chapter on 'Technical data'.
The unit is properly installed.
The unit is connected to the mains.
1. Turn the main switch A to position I.
2. Wait for one minute.
The LED (B) lights up on the Easypad.
The radiation indicator (C) lights up for approx. one second as a
functional check.
The home screen is displayed on the touchscreen for approximately
1 minute.
The program selection is then displayed on the touchscreen.
The forehead support and temple supports are completely open.
NOTE
The unit must not be switched on/off constantly.
Constant switching on and off reduces the service life of individual unit
components and results in increased power consumption.
➢ After switching the unit off, wait for approx. 60 seconds before
switching it on again.
5 Operation
5.1 Acquiring the X-ray image
49