CEREC AC Connect
With CEREC Omnicam
Operating Instructions (not valid for USA/Canada)
English

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CEREC AC Connect and is the answer not in the manual?

Questions and answers

Summary of Contents for Dentsply Sirona CEREC AC Connect

  • Page 1 07.2018 New as of: CEREC AC Connect With CEREC Omnicam Operating Instructions (not valid for USA/Canada) English...
  • Page 2: Table Of Contents

    Table of contents Dentsply Sirona Operating Instructions CEREC AC Connect Table of contents Dear Customer,....................... Contact information..................General data ......................Structure of the document................2.1.1 Identification of the danger levels ..........2.1.2 Formats and symbols used............Battery warranty................... Legend ......................
  • Page 3 Dentsply Sirona Table of contents Operating Instructions CEREC AC Connect Installation and startup.................... Transport and unpacking ................Disposal of packaging materials ..............Scope of supply ................... Initial startup....................6.4.1 Controls and functional elements........... 6.4.2 Operating state LED ..............6.4.3 Plug connections................
  • Page 4 Table of contents Dentsply Sirona Operating Instructions CEREC AC Connect 8.1.5.3 Removing the mirror sleeve.......... 8.1.5.4 Refitting the mirror sleeve..........8.1.5.5 Pre-cleaning of the mirror sleeve........8.1.5.6 Wipe disinfection for the camera and mirror sleeve ..8.1.5.7 High-level disinfection of the mirror sleeve....
  • Page 5: Dear Customer

    Dentsply Sirona 1 Dear Customer, Operating Instructions CEREC AC Connect 1.1 Contact information Dear Customer, ® Thank you for purchasing your CEREC AC Connect from Sirona. This unit enables you to create digital exposures and to transmit these to your dental laboratory.
  • Page 6: General Data

    2 General data Dentsply Sirona 2.1 Structure of the document Operating Instructions CEREC AC Connect General data Please read this document completely and follow the instructions exactly. You should always keep it within reach. Original language of the present document: German.
  • Page 7: Formats And Symbols Used

    Dentsply Sirona 2 General data Operating Instructions CEREC AC Connect 2.2 Battery warranty 2.1.2 Formats and symbols used The formats and symbols used in this document have the following meaning:  Prerequisite Prompts you to do something. 1. First action step 2.
  • Page 8: Legend

    2 General data Dentsply Sirona 2.3 Legend Operating Instructions CEREC AC Connect Legend Year of manufacture Safety labels Identifies labels/imprints on the unit (see Safety labels  [→ 14]). Product disposal symbol (see "Disposal  [→ 73]"). Storage battery pack disposal symbol (see "Disposal of the storage battery pack" ...
  • Page 9 Dentsply Sirona 2 General data Operating Instructions CEREC AC Connect 2.3 Legend Symbols on the packaging Take note of the following symbols on the packaging: Protect from moisture Fragile; handle with care Temperature during storage and transport Relative humidity during storage and transport...
  • Page 10: General Description

    3 General description Dentsply Sirona 3.1 Certification Operating Instructions CEREC AC Connect General description Certification CE mark This product bears the CE mark in accordance with the provisions of Council Directive 93/42/EEC of June 14, 1993 concerning medical devices. NOTE...
  • Page 11: Intended Use

    Dentsply Sirona 3 General description Operating Instructions CEREC AC Connect 3.2 Intended use Intended use This unit allows you to create digital impressions and transmit these to your dental laboratory. The unit may be operated only by medically trained and qualified personnel.
  • Page 12: Safety

    4 Safety Dentsply Sirona 4.1 Basic safety information Operating Instructions CEREC AC Connect Safety Basic safety information 4.1.1 Prerequisites NOTE Important information on building installation The building installation must be performed by a qualified expert in compliance with the national regulations. DIN VDE 0100-710 applies in Germany.
  • Page 13: Movement And Stability Of The Unit

    ● The maintenance and repair of this unit may be performed only by Dentsply Sirona or by agencies authorized by Dentsply Sirona. ● Components which have failed and influence the safety of the unit must be replaced with original (OEM) spare parts.
  • Page 14: Accessories

    In order to ensure product safety, this device may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. 4.1.7.1 Included accessories ●...
  • Page 15 Dentsply Sirona 4 Safety Operating Instructions CEREC AC Connect 4.2 Safety labels Plug connections of external interfaces CAUTION Additional devices connected to external interfaces must be tested according to the relevant standards, e.g.: EN 60601-1:1990 + A1:1993 + A2:1995 based on IEC 60601-1, EN...
  • Page 16 4 Safety Dentsply Sirona 4.2 Safety labels Operating Instructions CEREC AC Connect Heater plate CAUTION Risk of burns due to hot surface! ➢ Never touch the heater plate (A)! 64 32 087 D3492 D3492.201.07.24.02    07.2018...
  • Page 17: Electrostatic Charge

    Dentsply Sirona 4 Safety Operating Instructions CEREC AC Connect 4.3 Electrostatic charge Electrostatic charge 4.3.1 ESD warning labels ESD warning label CAUTION Risk of injury or damage to components from electrostatic discharge For electrical components labeled with an ESD warning label, observe the following instructions.
  • Page 18: About The Physics Of Electrostatic Charges

    4 Safety Dentsply Sirona 4.3 Electrostatic charge Operating Instructions CEREC AC Connect 4.3.3 About the physics of electrostatic charges What is an electrostatic charge? An electrostatic charge is a voltage field on and in an object (e.g. a human body) which is protected against conductance to ground potential by a nonconductive layer (e.g.
  • Page 19: Wireless Phone Interference With Equipment

    Dentsply Sirona 4 Safety Operating Instructions CEREC AC Connect 4.4 Wireless phone interference with equipment Wireless phone interference with equipment The use of mobile wireless phones in practice or hospital environments must be prohibited to ensure safe operation of the unit.
  • Page 20: Integration In A Network Or Connection To A Modem

    4 Safety Dentsply Sirona 4.5 Integration in a network or connection to a modem Operating Instructions CEREC AC Connect Integration in a network or connection to a modem NOTE Observe the following installation regulations The following installation regulations apply to integration of the...
  • Page 21: Technical Information

    Dentsply Sirona 5 Technical information Operating Instructions CEREC AC Connect 5.1 Technical description Technical information Technical description CAD system for high-precision intraoral optical impressions ● High-resolution, heated oral camera (3D camera) with removable reflective sleeve (reflective sleeve sterilizable with hot air) ●...
  • Page 22 5 Technical information Dentsply Sirona 5.1 Technical description Operating Instructions CEREC AC Connect PC hardware (V3.2.2) Special PC with the following equipment: ● Processor: Intel i7 5820K ● Memory: 2 x 8GB, 2133MHz DDR4-RAM ● DVD-R(W)/CD-R(W): Combined drive GH24NS ● Hard disk: 2 TB 2.5”...
  • Page 23: Technical Data

    Dentsply Sirona 5 Technical information Operating Instructions CEREC AC Connect 5.2 Technical data Technical data Type designation CEREC AC Acquisition unit Rated line voltage 100 - 240 VAC /50 - 60 Rated current 4.7 - 1.8 A Type of protection against electric shock...
  • Page 24: Electromagnetic Compatibility

    5 Technical information Dentsply Sirona 5.3 Electromagnetic compatibility Operating Instructions CEREC AC Connect Dimensions and weight Dimensions (WxHxD) in mm 350 x 1210 x 470mm in inches 13¾ x 47⅝ x 18½“ Weight ● without monitor and battery pack 38 kg (83.8 lbs) approx.
  • Page 25: Interference Immunity

    Dentsply Sirona 5 Technical information Operating Instructions CEREC AC Connect 5.3 Electromagnetic compatibility 5.3.2 Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
  • Page 26 5 Technical information Dentsply Sirona 5.3 Electromagnetic compatibility Operating Instructions CEREC AC Connect Interference immu- IEC 60601-1-2 Test level Compliance level Electromagnetic environment – nity tests guidelines Conducted RF in- d= [1.2] √P terference 150 kHz to 80 MHz IEC 61000-4-6...
  • Page 27: Working Clearances

    Dentsply Sirona 5 Technical information Operating Instructions CEREC AC Connect 5.3 Electromagnetic compatibility 5.3.3 Working clearances Recommended working clearances The UNIT is intended for operation in an electromagnetic environment, between portable and mobile RF where radiated RF interference is checked. The customer or the user of...
  • Page 28: Installation And Startup

    Operating Instructions CEREC AC Connect Installation and startup Transport and unpacking All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery. 1. Check the delivery note to ensure that the consignment is complete.
  • Page 29: Initial Startup

    Dentsply Sirona 6 Installation and startup Operating Instructions CEREC AC Connect 6.4 Initial startup Initial startup 6.4.1 Controls and functional elements Overview of the front panel Monitor ON/OFF switch Center trackball button Membrane keyboard Left trackball button CEREC Omnicam Trackball...
  • Page 30 6 Installation and startup Dentsply Sirona 6.4 Initial startup Operating Instructions CEREC AC Connect Operating state LED ON button Components of the Omnicam Press detent to release Camera windows Mirror sleeve Calibration set Sapphire glass (coated) NOTE CEREC Omnicam is calibrated The CEREC Omnicam is calibrated ex works (see "Calibrating...
  • Page 31 Dentsply Sirona 6 Installation and startup Operating Instructions CEREC AC Connect 6.4 Initial startup Overview of rear side Fuses Power connection Main switch USB port I = ON, 0 = OFF 64 32 087 D3492 D3492.201.07.24.02    07.2018...
  • Page 32: Operating State Led

    6 Installation and startup Dentsply Sirona 6.4 Initial startup Operating Instructions CEREC AC Connect 6.4.2 Operating state LED Operating state LED ON button LED not lit: Acquisition unit is switched off at main switch. LED lights up yel- Acquisition unit is switched on at main switch,...
  • Page 33: Plug Connections

    1. Connect the unit to the line voltage with the power cord. 2. Carefully insert the connector of the CEREC Omnicam cable into the coupling on the CEREC AC Connect, watching out for the guide nose. 3. Check the plug connections of the power supply and the camera.
  • Page 34: Insert Battery (Optional)

    6 Installation and startup Dentsply Sirona 6.4 Initial startup Operating Instructions CEREC AC Connect 6.4.4 Insert battery (optional) Bottom door Battery cover Storage battery Battery connector 1. Open the lower door on the back panel. NOTE Open with coin. Use a coin to open the latch. Turn counterclockwise.
  • Page 35: Changing From Right-Handed To Left-Handed Operation

    Ä The unit is dry and can be put into operation. CAUTION Use only the supplied power cord Use only the power cord supplied by Dentsply Sirona to connect the acquisition unit to the power supply. If the acquisition unit is switched on at the main switch, it can be activated with the ON button.
  • Page 36: Switching The Units Off

    6 Installation and startup Dentsply Sirona 6.4 Initial startup Operating Instructions CEREC AC Connect NOTE Possible data loss and PC malfunction: Switching the exposure unit off at the ON button during operation may cause data loss and PC malfunctions. ➢ Always switch the unit off as described in the chapter "Switching the units off".
  • Page 37: Battery-Backed Operation (Optional)

    At least every three months the CEREC AC Connect acquisition unit must be connected to the mains supply to charge the storage batteries.
  • Page 38 6 Installation and startup Dentsply Sirona 6.5 Battery-backed operation (optional) Operating Instructions CEREC AC Connect The color of the symbol has the following meaning: ● Green = line voltage present, fan functioning, temperature OK. ● Yellow = Unit running in battery-powered mode, all other operating parameters OK.
  • Page 39 Dentsply Sirona 6 Installation and startup Operating Instructions CEREC AC Connect 6.5 Battery-backed operation (optional) Monitoring window Explanation When the battery charge is almost depleted, the shut- down window opens. The operating system is then shut down after 30s and the PC can then be switched off.
  • Page 40: Operation

    7 Operation Dentsply Sirona 7.1 Camera warm-up time Operating Instructions CEREC AC Connect Operation Camera warm-up time When switching on the system, the camera needs to warm up for 15 - 20 minutes. If the coated sapphire glass of the Omnicam is not sufficiently warm, it may steam up during the acquisition.
  • Page 41: Taking Acquisitions With The Cerec Omnicam

    Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.3 Taking acquisitions with the CEREC Omnicam Taking acquisitions with the CEREC Omnicam CAUTION Hot surface! The coated sapphire layer of the CEREC Omnicam is preheated in the camera cradle. When removing the CEREC Omnicam from its cradle, the surface temperature of the mirror sleeve can be up to 51°C.
  • Page 42 7 Operation Dentsply Sirona 7.3 Taking acquisitions with the CEREC Omnicam Operating Instructions CEREC AC Connect WARNING Risk of injury for those diagnosed with epilepsy For persons who have been diagnosed with epilepsy, there is a risk of epileptic shock through the flashing light of the CEREC Omnicam.
  • Page 43: Directing The Camera

    Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.4 Directing the camera Directing the camera CAUTION After each use Prepare the camera once again after each patient. ➢ Follow the instructions in the "CEREC Omnicam  [→ 54]" section in order to avoid cross-contamination between patients.
  • Page 44: Occlusal Scan

    7 Operation Dentsply Sirona 7.4 Directing the camera Operating Instructions CEREC AC Connect 7.4.1 Occlusal scan Important: Ensure that the distance between the coated sapphire glass of the CEREC Omnicam and the scanned surface is observed. The distance must be between 0  -  15 mm (ideally: 5 mm). The camera does not rest on the teeth or the gums.
  • Page 45: Lingual Scan

    Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.4 Directing the camera 7.4.3 Lingual scan ü The CEREC Omnicam is on the tooth that is positioned next to the preparation in the distal direction. 1. Rotate the CEREC Omnicam from 90° in the buccal direction to around 45°...
  • Page 46: Buccal Registration

    7 Operation Dentsply Sirona 7.4 Directing the camera Operating Instructions CEREC AC Connect 7.4.5 Buccal registration A buccal registration can be used to establish contact with the antagonist. ü The jaw with the preparation is scanned. 1. Scan the occlusal, buccal and lingual view of the antagonist (see the section “Occlusal scan ...
  • Page 47: Scanning The Quadrant And Jaw

    Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.4 Directing the camera 7.4.7 Scanning the quadrant and jaw The following scan regulation applies for the acquisition of a complete quadrant or jaw arch. The first (fourth) quadrant is scanned up to the opposite second front tooth by moving the camera in parallel along the jaw arch.
  • Page 48 7 Operation Dentsply Sirona 7.4 Directing the camera Operating Instructions CEREC AC Connect Completing the scanning process 1. Start as indicated above, on the occlusal surface of the right terminal tooth, and scan it occlusally. Tilt the camera by 45° in a palatinal direction (oral) and guide it from the distal to the mesial.
  • Page 49 Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.4 Directing the camera The following scan sequence is implemented for the opposite second (third) quadrant: 1. Start by placing the device on the occlusal surface of a premolar, that has already been scanned, and guide the camera palatally (orally) at a mesial tilt of up to 90°...
  • Page 50: Completing Measurements

    7 Operation Dentsply Sirona 7.4 Directing the camera Operating Instructions CEREC AC Connect ● When performing this activity, be careful not to accidentally cut out any areas that e.g. are located behind the model or are otherwise cut away from the line.
  • Page 51: Software For The Cerec Omnicam

    Dentsply Sirona 7 Operation Operating Instructions CEREC AC Connect 7.5 Software for the CEREC Omnicam Software for the CEREC Omnicam For more information on using the CEREC SW, refer to the Operator's Manual. You can find additional information on setting the CEREC Omnicam in the "Adjusting the CEREC Omnicam ...
  • Page 52: Maintenance

    Some countries have legal regulations which require regular safety inspections of electrical devices or systems by the operator. Dentsply Sirona would like to draw your attention to the fact that a so- called "retest" (repeat test) must be carried out for the CEREC AC acquisition unit at least every three years.
  • Page 53: Monitor Screen

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.1 Care, cleaning, disinfection, and sterilization 8.1.2 Monitor screen Cleaning NOTE Never spray the monitor screen with disinfectant or cleaning agents! The monitor screen must be wiped with a soft cloth.
  • Page 54: Trackball Holder

    8 Maintenance Dentsply Sirona 8.1 Care, cleaning, disinfection, and sterilization Operating Instructions CEREC AC Connect 8.1.4 Trackball holder 1. Rotate the cover ring counterclockwise and remove it. 2. Clean inner surface of cover ring (A) with ethanol (commercially available cleaning alcohol).
  • Page 55: General Information

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.1 Care, cleaning, disinfection, and sterilization 8.1.5.1 General information The CEREC Omnicam is a very sensitive optical device and must therefore be handled with the utmost care. Protect the coated sapphire...
  • Page 56: Refitting The Mirror Sleeve

    8 Maintenance Dentsply Sirona 8.1 Care, cleaning, disinfection, and sterilization Operating Instructions CEREC AC Connect 8.1.5.4 Refitting the mirror sleeve ➢ Carefully refit the mirror sleeve until it locks in place. NOTE There is a risk of damaging the camera window or the coated sapphire glass if the mirror sleeve is not pushed straight toward the front.
  • Page 57: High-Level Disinfection Of The Mirror Sleeve

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.1 Care, cleaning, disinfection, and sterilization 8.1.5.7 High-level disinfection of the mirror sleeve NOTE The hot air sterilization and high-level disinfection must not be combined. The complete process for high-level disinfection (HLD) is as follows –...
  • Page 58 8 Maintenance Dentsply Sirona 8.1 Care, cleaning, disinfection, and sterilization Operating Instructions CEREC AC Connect 6. Push the plug completely into the metal bracket. 7. Press the mirror sleeve onto the plug while holding the metal bracket firmly in place in order to ensure that no particle contamination or fluids penetrate the inside of the mirror sleeve.
  • Page 59 Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.1 Care, cleaning, disinfection, and sterilization 9. Insert the metal bracket with the metal sleeve. 10. Place the cover on the container and leave the mirror sleeve in the ® disinfectant for at least 12 minutes (CIDEX OPA).
  • Page 60 8 Maintenance Dentsply Sirona 8.1 Care, cleaning, disinfection, and sterilization Operating Instructions CEREC AC Connect 11. Remove the cover and take the bracket with the mirror sleeve out of the container. 12. Carefully rinse the mirror sleeve with water from the mains while holding it with the bracket.
  • Page 61: Hot Air Sterilization Of The Mirror Sleeve

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.1 Care, cleaning, disinfection, and sterilization 17. Dry the entire outer surface of the mirror sleeve as well as the inner section using a soft lint-free cloth. Ensure that no fluid is able to penetrate the inside of the mirror sleeve.
  • Page 62: Use Of Cerec Omnicam Disposable Sleeves

    8 Maintenance Dentsply Sirona 8.1 Care, cleaning, disinfection, and sterilization Operating Instructions CEREC AC Connect 8.1.5.9 Use of CEREC Omnicam disposable sleeves The CEREC Omnicam can be operated with disposable plastic sleeves in order to ensure maximum infection control. The sleeves are available using order number 66 32 264.
  • Page 63: Calibrating Cerec Omnicam

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.2 Calibrating CEREC Omnicam Calibrating CEREC Omnicam The measurement procedure used by the system requires the use of a calibrated CEREC Omnicam. The CEREC Omnicam is factory- calibrated. Then calibrate the CEREC Omnicam after every reinstallation and after every transport.
  • Page 64 8 Maintenance Dentsply Sirona 8.2 Calibrating CEREC Omnicam Operating Instructions CEREC AC Connect Calibrate the camera 1. Remove the protective cap from the calibration set. 2. Mount the calibration set on the tip of the camera until it locks into place.
  • Page 65 Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.2 Calibrating CEREC Omnicam Measuring the position of the exit window 1. Mount the bottom side of the calibration set to the tip of the camera. 2. Click on the "OK" button.
  • Page 66: Color Calibration

    8 Maintenance Dentsply Sirona 8.3 Color calibration Operating Instructions CEREC AC Connect Color calibration General information NOTE Faulty color analysis The color analysis can be negatively impacted due to strong light incidence and it can lead to varying results. ➢ Set the model CEREC Omnicam up so that it is not located directly in the beam path of an extreme light source (e.g., the...
  • Page 67 Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.3 Color calibration Carry out a color calibration also after changing a mirror sleeve. Heavily scratched mirror sleeves may not be used for a color analysis. Storing a color-calibration set The color-calibration set must be stored in its packaging in a dry place which is protected from light.
  • Page 68 8 Maintenance Dentsply Sirona 8.3 Color calibration Operating Instructions CEREC AC Connect Color-calibrating the camera 1. Make sure that the CEREC Omnicam is clean, disinfected and dry. 2. Remove the color-calibration set from the packaging. 3. Use the CEREC Omnicam to scan the QR code on the underside of your color-calibration set.
  • Page 69 Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.3 Color calibration Replacing the color calibration set Please note that the color calibration set ● can only be used with CEREC SW software ≥ 4.5 or Sirona Connect SW ≥ 4.5 or CEREC Premium SW ≥ 4.5.
  • Page 70: Replacing The Main Fuse

    8 Maintenance Dentsply Sirona 8.4 Replacing the main fuse Operating Instructions CEREC AC Connect Replacing the main fuse WARNING Electric shock Disconnect the power plug at the unit end before replacing the fuses. NOTE Fuse type Use only fuses of the same type in the fuse module!
  • Page 71: Charge Battery (Optional)

    Dentsply Sirona 8 Maintenance Operating Instructions CEREC AC Connect 8.5 Charge battery (optional) Charge battery (optional) NOTE Reduced buffer cycles After around 1000 buffer cycles the capacity of the battery fades due to the nature of the battery technology used.
  • Page 72: Replace Battery (Optional)

    8 Maintenance Dentsply Sirona 8.6 Replace battery (optional) Operating Instructions CEREC AC Connect Replace battery (optional) Bottom door Battery cover Storage battery Battery connector 1. Open the lower door on the back panel. NOTE Open with coin. Use a coin to open the latch. Turn counterclockwise.
  • Page 73: Disposal

    Dentsply Sirona 9 Disposal Operating Instructions CEREC AC Connect 8.6 Replace battery (optional) Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
  • Page 74: Disposal Of The Storage Battery Pack

    9 Disposal Dentsply Sirona 9.1 Disposal of the storage battery pack Operating Instructions CEREC AC Connect Disposal of the storage battery pack The storage battery pack must be subjected to recycling if it becomes defective or reaches the end of its service life. Recycling is performed via Sirona.
  • Page 75: Pc Diagnostic Tool

    Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.1 Checking the BIOS configuration (only V2.2.1 - V2.2.3) PC Diagnostic Tool Checking the BIOS configuration (only V2.2.1 - 10.1 V2.2.3) Check the BIOS configuration before using the PC diagnostic tool. If the configuration is set incorrectly, the PC diagnostic tool will not recognize the hard drive and cannot check it.
  • Page 76: Start Diagnostic Tool

    10 PC Diagnostic Tool Dentsply Sirona 10.2 Start diagnostic tool Operating Instructions CEREC AC Connect 4. Set the entry "ASM1061 Storage Controller" to "IDE Mode". 5. In the menu go to "Exit" and "Save Changes & Reset" or press the "F10" button to save the settings.
  • Page 77: Test Procedure

    Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.3 Test procedure 6. Select the test using the arrow keys. 7. Confirm your selection with the "Return" key on the keyboard. Ä The test starts. Test procedure 10.3 10.3.1...
  • Page 78 10 PC Diagnostic Tool Dentsply Sirona 10.3 Test procedure Operating Instructions CEREC AC Connect Ä The test starts. The entire test run takes approx. 30 minutes. Ä The result, i.e. "Pass" or "Fail", appears at the end of the test (see sections entitled "Test result: Pass" or "Test result: Fail").
  • Page 79: Technician Diagnostics

    Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.3 Test procedure Test result: Fail An error was found on the PC-specific hardware. Replacing the PC component or the PC may be required. An image restore is not advisable.
  • Page 80 10 PC Diagnostic Tool Dentsply Sirona 10.3 Test procedure Operating Instructions CEREC AC Connect Ä The test starts. The entire test run takes approx. 30 minutes. Ä The result, i.e. "Pass" or "Fail", appears at the end of the test (see sections entitled "Test result: Pass" or "Test result: Fail").
  • Page 81: Sirona Windows Diagnostics

    Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.3 Test procedure Test result: Fail An error was found on the PC-specific hardware. Replacing the PC component or the PC may be required. An image restore is not advisable.
  • Page 82: Troubleshooting

    10 PC Diagnostic Tool Dentsply Sirona 10.4 Troubleshooting Operating Instructions CEREC AC Connect ➢ Check the result in the "Windows Stress Test" window. Tip: You may have to rearrange the windows to be able to see the "Windows Stress Test" window.
  • Page 83: Customer Diagnostics & Technician Diagnostics

    Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.4 Troubleshooting 10.4.1 Customer Diagnostics & Technician Diagnostics Test Test description / condition Action if problems occur Display Script Control command without a test n.a. function Activity is rotating cursor...
  • Page 84 10 PC Diagnostic Tool Dentsply Sirona 10.4 Troubleshooting Operating Instructions CEREC AC Connect Test Test description / condition Action if problems occur Test WDC___TYP HDD___Short A range of destruction-free tests on 1. For the CEREC AC check the Self-Test the hard drives installed in a system setting of the ASM1061 Storage are conducted with these functions.
  • Page 85 Dentsply Sirona 10 PC Diagnostic Tool Operating Instructions CEREC AC Connect 10.4 Troubleshooting Test Test description / condition Action if problems occur Test VGA RAM This test checks the graphics mem- ➢ Replace the graphics card. ory installed in the graphics card that is currently active.
  • Page 86: Sirona Windows Diagnostics

    10 PC Diagnostic Tool Dentsply Sirona 10.4 Troubleshooting Operating Instructions CEREC AC Connect 10.4.2 Sirona Windows Diagnostics Test Test Description Action if problems occur Processor(s) Continuous load and inspection of pro- ➢ Replace the PC. cessor cores Motherboard testing Continuous load and inspection of the ➢...
  • Page 87: Appendix

    Dentsply Sirona 11 Appendix Operating Instructions CEREC AC Connect 11.1 DVD playback Appendix DVD playback 11.1 DVD videos can be played back on the acquisition unit via "Windows Media Center" . ➢ Start the program via the corresponding icon or via "Start" / "All Programs"...
  • Page 88: Seal On Pc Slide-In Module

    11 Appendix Dentsply Sirona 11.3 Seal on PC slide-in module Operating Instructions CEREC AC Connect Seal on PC slide-in module 11.3 NOTE If the seal is broken, all warranty claims regarding the PC slide- module automatically expire. The PC slide-in module may be opened only by an authorized dental technician.
  • Page 89: Index

    Dentsply Sirona Index Operating Instructions CEREC AC Connect Index Acquisition unit Foot pedal, 29 Overview, 29 Fuse Air pressure Fuse type, 70 Operation, 23 Order No., 70 Storage, 23 Replacement, 70 Transport, 23 Fuses, 31 Ambient temperature Operation, 23 Heater plate, 29 Application, 11 HUB, 20 Building installation, 12 Insertion Trackball, 34 Intended use, 11 Calibration set, 30 Camera Warm-up time, 40...
  • Page 90 Index Dentsply Sirona Operating Instructions CEREC AC Connect Plug connections, 33 Power connection, 31 Voltage selection insert, 70 Product safety, 14 Protection against medicaments, 53 Water, 23 Protection class, 23 Weight, 24 Protective ground wire, 20 Rated current, 23 Rated line voltage, 23 Relative humidity Operation, 23 Storage, 23 Transport, 23 Repair, 13 Safety instructions, 6 Sapphire glass, 30, 55...
  • Page 92 We reserve the right to make any alterations which may be required due to technical improvements. © Sirona Dental Systems GmbH Sprache: englisch Printed in Germany D3492.201.07.24.02      07.2018 Ä.-Nr.:  125 755 Sirona Dental Systems GmbH 64 32 087 D3492 Fabrikstr. 31 Order No   ...

Table of Contents