Operation

CAUTION
Risk of injury on calibration pins/instruments
If you reach into the processing chamber (e.g.: when inserting/
removing a ceramic block, changing instruments, or inserting/
removing a calibration phantom), you may injure your hand on the
calibration pins/instruments.
Be careful not to brush against the calibration pins/instruments with
your hand.
Always insert your hand in the processing chamber underneath the
calibration pins/instruments.

Configuration (CEREC MC/MC X)

6.1
In the "Devices" area of the CEREC SW software, various settings can
be subsequently modified.
1. Click the "Configuration" button in the system menu.
2. Click on the "Devices" button.
3. Click on the unit that you wish to configure.
Bar code reader
If a bar code reader is used, e. g. for in inCoris ZI, the corresponding
box must be activated. The bar code reader will then always be used to
read a bar code.
Extraction Unit
If a suction device is connected, and the check mark is set, the dry
milling process is automatically started when processing zirconium
oxide. To deactivate dry milling, you can remove the check mark once
again.
6 Operation
6.1 Configuration (CEREC MC/MC X)
35