5 Installation and startup
5.4 Commissioning
28
2. Connect the plug-in power supply included in the scope of delivery
with the wireless module HW 8614/F2 and plug it into the power
supply.
3. If necessary, secure the wireless module in the selected operating
position using the preassembled Velcro ® tape.
In doing so, ensure that the rod antenna is vertically positioned.
4. Pair the wireless module of the milling and grinding unit as
described in the installation instructions included with the wireless
module (REF 62 80 064).
You can pair multiple milling and grinding units with a CEREC
acquisition unit. If more than 2 milling and grinding units are operated at
once, the limited data bandwidth may cause processing time delays.

Dentsply Sirona

Operating Instructions CEREC MC / MC X
63 98 643 D3439
D3439.201.07.11.02    01.2018