Information On The Seal Of Approval; Entering The Bar Code - Dentsply Sirona CEREC MC Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC MC / MC X
63 98 643 D3439
D3439.201.07.11.02    01.2018
6.4.5

Information on the seal of approval

Proper selection and processing of the material are decisive for the
long-term clinical success of the restoration, especially in the case of
zirconia. However, different types of zirconia require individually
matched machine parameters. This is the reason why you can and must
select different types of zirconia in the software. These machine
parameters are coordinated between material partners in complex
development processes. In addition to the desired fit and surface
quality, they also guarantee a maximum degree of material and
equipment safety. Consistently high quality of the grinding or milling
result and the fit can be guaranteed and damage to the processing
machines can be excluded only if certified materials are used.
NOTE
Block without seal of approval
If a block is found without a seal of approval during the machining
process (milling or grinding), the following message appears:
"No quality label was recognized on the block. The grinding and
milling processes as well as the instruments are specially verified for
certified materials. Certified materials can be identified by the
engraved "inLab" lettering on the block. The use of zirconium oxide
materials without quality label can lead to inferior results as well as
increased wear on the device and instruments.
Do you still wish to start the manufacturing process?"

Entering the bar code

6.5
Barcode Reader active
If you have activated the option "Barcode Reader" in the system
configuration dialog (e.g. for CEREC Zirconia), you must read-in both
bar codes with the bar code reader. To do this, hold the bar code reader
tilted to a slight angle and move it over both of the bar codes on the
block continuously and evenly.
If the reading process fails, you can read-in the bar code once again by
pressing "Retry" (button 1 on the unit display). Alternatively, you also
can enter the substitute code (8-digit character string, e.g. *1234XYZ)
on the PC manually.
No bar code reader available
➢ Enter the substitute code (8-digit character string, e.g. *1234XYZ)
on the PC manually.