Water Tank Cerec Mc Xl Basic - Dentsply Sirona CEREC MC Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC MC / MC X
63 98 643 D3439
D3439.201.07.11.02    01.2018
7.3.2.2

Water tank CEREC MC XL Basic

A
Water tank
A
Filter insert
B
Tank cap
To change the water, proceed as follows:
ü The unit is switched on.
ü No machining process is running.
1. Pull out the water tank at the front of the unit.
2. Open the drain opening.
3. Drain the water tank.
4. Turn the tank cap counter-clockwise and take it off. If you find the
tank cap hard to open by hand, use the tank cap opener (see
"Opening the tank cap").
5. Unscrew the side cap.
6. Remove the filter insert from the tank and clean the filter thoroughly
under running water.
7. Rinse the water tank.
8. Insert the cleaned filter with its cap into the unit and screw it tight.
9. Close the drain opening.
NOTE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not
permitted.
Do not use any cleaning agents.
10. Add approx. 75 ml of DENTATEC to the tank.
11. Fill the tank with water until the filter insert is completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
12. Wait for a short time until the filter insert is completely soaked; then
add an appropriate amount of water.
B
D
C
C
Tank
D
Tank drain
7 Service
7.3 Changing the water
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerec mc x

Table of Contents