Dentsply Sirona
Operating Instructions CEREC MC / MC X
63 98 643 D3439
D3439.201.07.11.02 01.2018
Changing the filter
7.8
NOTE
Change the filter regularly!
Clean the filter regularly and change it immediately when damaged.
Otherwise, change it every 3 months.
If a message appears stating that the water pressure is too low, you
must clean the filter or, if it is damaged, replace it immediately.
CAUTION
Filter
Use only filters approved by Dentsply Sirona!
7.8.1
Water tank MC / MC X
B
A
Water tank
A
Filter insert
B
Tank cap
ü The tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off. If you find the
tank cap hard to open by hand, use the tank cap opener.
3. Take the filter insert out of the tank.
4. Rinse the water tank.
7.8 Changing the filter
C
Tank
D
Tank drain
7 Service
C
D
59